Pular para o conteúdo principal
GE Healthcare · CTLightSpeed18261418260751200

LightSpeed 2.X (Plus, QX/i) - Octane

Symptom

Gantry tilts backward but will not tilt forward or tilts slowly Gantry tilts backward but will not tilt forward or tilts forward slowly

Error message

Error: 182614 Tilt drive is not stopping within 1500 msec. Expected stopping position XX (.01 deg) Current position YY (.01 deg). See Long Description for error details and example Host: ETC Process: Tilt File: tmcShutDown.c 1.31 1/29/93 Line: 121 Function: PATIENT ANATOMY POSITIONING : Elevation - Tilt Scan: 5478/4/0 Type: None/Unknown Exception Level: Pri/Soft Time: 16:25:56:115 Log Series: 151 HW/SW error: Tilt drive is NOT stopping within 1500 msec. Expected Stopping Position -50 (.01 deg). Current Position -23 (.01 deg). Tilt Direction: 0 (0 = BCK 1 = FWD). Error: 182607 Host: ETC Process: Tilt File: CmdMotion.c 1.21 2/17/95 Line: 139 Function: PATIENT ANATOMY POSITIONING : Elevation - Tilt Scan: 7839/2/0 Type: None/Unknown Exception Level: Pri/Soft Time: 12:12:01:160 Log Series: 617 HW/SW error: Velocity tracking error. Actual Velocity 0 (.01 deg/sec) Expected Velocity 100 (.01 deg/sec) Tol 80 (.01 deg/sec) Cur Pos -2181 (.01 deg) Tilt Dir: 1 (0 = BCK 1 = FWD). INFO: Probable Problem could be 180V Power is LOW.

Error: 182607 HW/SW error: Tilt Velocity tracking See LD for more details Thu May 11 23:04:54 2006 Error: 182607 Host: ETC Process: Tilt File: CmdMotion.c 1.21 2/17/9

Applies to

Models
LightSpeed
Product
LightSpeed 2.X (Plus, QX/i) - Octane Always S-DAS 2.4.2_H2.4M5 LightSpeed 2.X (Plus, QX/i) - Octane (Pegasus or Rhapsode IG) - Always S-DAS Method of visual identification Method of remote or software identification - See attachment below for a complete list of software commands and typical results. check_config List of Supported Console Types Octane with Pegasus Recon hardware. List of softwares the product can run 2.4.2_H2.4M5 FAQ's Table weight limit - Up to 400 lb. (180 kg) with ± .25 mm positional accuracy. 450Ib (205 kg) max weight Tube type - D3186T Current production details - See Technical Reference Manual found on the Common Document Library Tips and workarounds, where to find it - Common Document Library(CDL) . Search for the latest FMI software release. In the furnished materials will be the list of various manuals with part numbers for the various languages. Example: English versions 2211223-100 FEATURES/OPTIONS STD/OPTION Cat # Digital Laser Camera I/F B7700PD Smart Helical (s/w only) Standard B7530TK Fast Recon (s/w only) Option B7700JF BMD (requires B7710MP or MN) Option B7710MK Smart Prep Option B7530SP Smart Score ? ? 3D Analysis (Requires D3D) Option B7540TD Navigator (s/w only) Option B7540N DentaScan (s/w only) Option B7540SD Direct 3D (Requires B7710SA) Option B7810SS HIS/RIS Option (h/w and s/w) Option B7500PN Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
46-264368G3 ETC board See long description for details Part #: 46-264368G3 Part Type: GP CONSUMABLE - RETURNABLE Part Weight: 0.00 LBR Repairable: No Returnable: Yes Hazardous Material: N Description: ENHANCED TABLE CONTROL HELIOS 2 FRU: Yes Status: Orderable Comments: Not Available
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.