Pular para o conteúdo principal
GE Healthcare · CTDiscoveryVCT2601331915001826013319350179

"Scan hardware Stopped Scan" with collimator errors.

Symptom

"Scan hardware Stopped Scan" with collimator errors. "Scan hardware Stopped Scan" with collimator errors

Error message

Error: 260133191 Filter A / B Switch Error. Both Contacts Are Either Open or Closed. See long description for full error Error: 260133191 Filter A or B Switch Error: Both Contacts are either Open or Closed. Tue Jan 17 07:24:14 2006 Error: 260133191 orp orp_subsys filter.cxx 81 Host : Collimator Ermes # : 260133191 Exception Class : Secondary Severity : Pri/Soft File : filter.cxx Line# : 81 Function : No System Function Reported Scan Type : Static Scan: 50018/1/1 Filter A or B Switch Error. Both Contacts Are Either Open or Closed. They Should be the inverse of each other Filter Switch NO Contact: Open Filter Switch NC Contact: Open

Error: 260133193 - Filter A failed to Home (or Filter B failed to Home) See long description for complete error code Sun Mar 18 05:49:17 2007 Error: 260133193 Host: orp Process: orp_subsys File: filter_home.cxx Line: 158 Sun Mar 18 05:49:16 2007 Host : Collimator Ermes # : 260133193 Exception Class : Abort Severity : Primary File : filter_home.cxx Line# : 158 Function : No System Function Reported Scan Type : None/Unknown Scan: 50179/7/0 Filter A failed to Home. OR it could be filter A and B. Sun Mar 18 05:49:17 2007 Error: 260133193 Host: orp Process: o

Applies to

Models
Discovery, VCT
Product
Discovery CT750 HD V40_PS_HD64_G_GTL V40_PS_HD64_G_GTM Discovery HD CT750 Method of visual identification HD gantry: Gray top and Gray Side covers with HD Markings Console Type: GOC6 has black SCIM and tower face Method of remote or software identification - See attachment below for a complete list of software commands and typical results. check_config swhwinfo “swhwinfo” will have the following new switches available · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) -h or –help = Outputs complete details about all the available switches associated with swhwinfo swhwinfo –c –m –o -da List of softwares the product can run (Consult attached FCT Software Cross Reference): 08MW39.4 - M3 (see attachement for M3 available options) 09MW29.14 - M4 (HD Options = M3 plus Gemstone Spectral Imaging, Gemstone Spectral Review and volume Helical Shuttle ) Example: Of difference with options software [insite@ct01 ~]$ swokinstall -p Product Software Option Keys: Gemstone Spectral Imaging Permanent HelicalShuttle Permanent Xtream Recon Permanent FAQ's Table weight limit: 500 lbs Tube Type - D3702T Green D3701T Current production - M4 Refer to product data sheets or Technical reference Manual, English 5317222-1EN Revision : 2 or later Refer to Application Tips and Work-Arounds Direction 5311815-1EN Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
5410597 (replaces 5222001, 5222001-2 & 5222001-500 and 64 slice 5130001) VCT/HD Common Collimator FRU - used on 64 slice VCT and HD - see SN11091210a Part #: 5410597 Part Type: GP CONSUMABLES Part Weight: 0.00 LBR Repairable: No Returnable: No Hazardous Material: N Description: VCT VCTHD Common Collimator FRU-ROHS FRU: Yes Status: Orderable Comments: **FE ASSIST** 5410597 replaces 5222001-500, to be used on Gantries 5124069-x and 5232083-x. For Gantry 5126093-x must use Collimator 5357424. Refer to SN11091210a (212001261 Sept 22,2011) Old FRU Compatible No Comments Per ECR 2120523, 212008898, 30 Aug 2011. Additional FRU Needed No Old FRU Scrapped No Old FRU Upgradable No Use on Hands Yes Use Demand History Yes Ultimate Prime 5410597 New FRU Backward Compatible Yes Old FRU Usable Yes Change Type Direct ECO Number ECR2120523 Part #: 5222001 Part Type: GP CONSUMABLE - RETURNABLE Part Weight: 0.00 LBR Repairable: No Returnable: Yes Hazardous Material: N Description: VCTHD Collimator Asm with Rating Plate FRU: Yes Status: Orderable
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required

Related GE Healthcare CT issues