GE Healthcare · CTLightSpeed210000434
LightSpeed RT16 & RT16 Xtra and Vision
Symptom
Scanner failed to scan with Image Allocation errors See Long Description for error examples Scanner failed to scan with Error: 2021 mkdir failed /usr/g/sdc_image_pool Input/output error and Error: 210000434 IMAGE ALLOCATION has returned null linkedTuple or null linkedTuple->data.
Error message
Error: 2106 Image Storage Full See Long Description for complete error Wed Aug 30 16:07:53 2006 Error: 2106 Host: IRIS Process: imserver File: imsprim.c Line: 742 Image Storage Full Error: 210000434 IMAGE ALLOCATION has returned null linkedTuple or null linkedTuple->data See Long Description for complete error Fri Apr 21 07:41:23 2006 Error: 210000434 Host: lightspeed Process: dms File: SaveContext.cxx Line: 184 I MAGE ALLOCATION has returned null linkedTuple or null linkedTuple->data.
Applies to
- Models
- LightSpeed
- Product
- LightSpeed RT16 & RT16 Xtra and Vision WB_H_G16_G_GTL = RT16, Xtra WB_H_G16_G_GTL650 = Vision WB_H_G16_G_GTM = RT16, Xtra or Vision LightSpeed RT16 & RT16 Xtra with GRE and GDAS16 gantry Links to the Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.