GE Healthcare · MRSigna22494321011620

Scan stops with PCI erorr 2249432 Harrier Err Enable Reg due to defective MGD power supply.

Symptom

Intermittent scan stops with PCI error 2249432 Harrier Err Enable TPS clears issue, but reoccurs within 1 hour

Error message

2249432 A PCI error occurred. The MCP820 received a Target Abort. See Long Description Error: 2249432 Host: EMR1 Process: diagCoord File: excLog.c Line: 1389 Task ID: 0x6eca5f0 Time: 1011620 Object: AGP Exception Class: Unknown Function: Software/Firmware/Config: N/A A PCI error occurred. The MCP820 received a Target Abort. Harrier Err Enable Reg: 0xFFF0E000; Harrier Err Status Reg: 0x40000000. Harrier Err Address Reg: 0x09016034; Harrier Err Attribute Reg: 0x0600. Additional information in /usr/g/service/log/vxPciLg-f6f77 and /usr/g/service/log/vxPciSum-f6fa4

Applies to

Models
Signa
Product
Signa HDxT This product is for 1.5T Signa HDxT systems with 15.0 software and HDi systems with 15.0 software. This product is for 1.5T Signa HDxT systems with 15.0 software and HDi systems with 15.0 software.
Software
15.0 For Product Signa HDxT and HDi only. -- NOT to be used with HDe. -- For Product Signa HDxT and HDi only. -- NOT to be used with HDe. --
Component
5181495-3 400W Quad Output 6Ux8HP cPCI Power Supply This is a replacement for EOPL 2294300-8
Area
MR

Full cause & step-by-step resolution