Attention The detector is not at the proper temperature.
Symptom
Attention The detector is not at the proper temperature. see long description for more detail. Attention The detector is not at the proper temperature. If you wish to proceed with this scan select the "OK" button, however IQ may be degraded including image artifacts. Please contact GE Service
Error message
Error: 260132914 Rail Temperature Out Of Range See Long Description for complete error Temperatures vary. Wed Jun 21 14:51:53 2006 Error: 260132914 Host: orp Process: orp_subsys File: ThermalManager.cxx Line: 2071 Wed Jun 21 14:51:52 2006 Host : Das Ermes # : 260132914 Exception Class : Secondary Severity : Sec/Soft File : ThermalManager.cxx Line# : 2071 Function : No System Function Reported Scan Type : None/Unknown Scan: 81/201/0 Rail Temperature Out Of Range Sensor: Rail Low Channel Temperature: 3640 <--- In this case, temp was always too LOW Average A/D Temperature: 0 Limits: 3650, 3950 Error: 244653 Host Querying Detector Temperature Status See LD for complete error Sun Jun 25 13:08:53 2006 Error: 244653 Host: gect02 Process: fwmgr File: fwmgr_events.c Line: 11634 Function: No system function reported Scan Type: Not scanning Scan: 0/0/0 Scan Seq Id: 0 Host Querying Detector Temperature Status: StateFlag:2 curTe mp:3180, leftTemp:3620 rightTemp:3730 Time:0 Fri Feb 25 07:51:11 2011 Error: 244653 Host: MT_CT2 Process: fwmgr File: fwmgr_events.c Line: 11915 Function: No system function reported Scan Type: Not scanning Scan: 0/0/0 Scan Seq Id: 0 Host Querying Detector Temperature
Applies to
- Models
- Discovery, VCT
- Product
- Discovery CT750 HD V40_PS_HD64_G_GTL V40_PS_HD64_G_GTM Discovery HD CT750 Method of visual identification HD gantry: Gray top and Gray Side covers with HD Markings Console Type: GOC6 has black SCIM and tower face Method of remote or software identification - See attachment below for a complete list of software commands and typical results. check_config swhwinfo “swhwinfo” will have the following new switches available · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) -h or –help = Outputs complete details about all the available switches associated with swhwinfo swhwinfo –c –m –o -da List of softwares the product can run (Consult attached FCT Software Cross Reference): 08MW39.4 - M3 (see attachement for M3 available options) 09MW29.14 - M4 (HD Options = M3 plus Gemstone Spectral Imaging, Gemstone Spectral Review and volume Helical Shuttle ) Example: Of difference with options software [insite@ct01 ~]$ swokinstall -p Product Software Option Keys: Gemstone Spectral Imaging Permanent HelicalShuttle Permanent Xtream Recon Permanent FAQ's Table weight limit: 500 lbs Tube Type - D3702T Green D3701T Current production - M4 Refer to product data sheets or Technical reference Manual, English 5317222-1EN Revision : 2 or later Refer to Application Tips and Work-Arounds Direction 5311815-1EN Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Component
- 5199438 Stationary CFC Assembly - VCT / HD see Long Description 5199438 Stationary CFC Common Fan Controller Asm VCT / HD Part #: 5199438 Part Type: GP CONSUMABLE - RETURNABLE Part Weight: 0.00 LBR Repairable: No Returnable: Yes Hazardous Material: N Description: CFC Assembly-VCT HD FRU: Yes Status: Orderable
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.