GE Healthcare · CTLightSpeed26000742226000740250444
LightSpeed 4.X - GRE
Symptom
Scanner Hardware Stopped Scan with DAS parity errors
Error message
Error: 260007422 Data Parity Error detected in the 5B Data Stream on Converter Board 2 or the DCB Board. See long description Mon Feb 1 06:34:22 2010 Error: 260007422 Host: DAS Process: DCB Firmware File: parity_16.c Line: 247 Function: DATA ACQUISITION : Scan Data Acquisition Scan: 0/0/0 Type: None/Unknown Exception Level: Pri/Soft Time: 06:34:23:235 Log Series: 31 SCAN ABORT -- DAS Has Detected An Error.^B Data Parity Error detected in the 5B Data Stream on Converter Board 2 or the DCB Board. (CSI:1 RSM:0x1303). Error: 260007402 Converter Board(s) Asserted a Board FAULT. See Long Description for complete error Tue Dec 28 09:41:31 2004 Error: 260007402 Host: orp Process: orp_subsys File: cnvmgr.c Line: 605 Tue Dec 28 09:41:30 2004 Host : Das Ermes # : 260007402 Exception Class : Abort Severity : Secondary File : cnvmgr.c Line# : 605 Function : No System Function Reported Scan Type : Static Scan: 50444/1/1 Converter Board(s) Asserted a Board FAULT.
Applies to
- Models
- LightSpeed
- Product
- LightSpeed 4.X - GRE See Long Description for detailed information - H2_P_M16 Product Name: LightSpeed 4.X with GRE Product Photographs GE HEALTHCARE Method of remote or software identification - See attachment below for a complete list of software commands and typical results 1. check_config 2 . swhwinfo “swhwinfo” will have the following switches: · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) -h or –help = Outputs complete details about all the available switches associated with swhwinfo swhwinfo –c –m –o -da List of Supported Console types: GOC3 (GRE) GOC4 (GRE) List of software the product can run: 07MW11.10 SP2.0, SP3.0, SP3.1 FAQ's Table weight limit - 180 kg (400 lbs) normal operation - 205 kg (450 lbs) maximum allowed with decreased accuracy Tube type - D3186T D3186T(green) Current production - M4 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Component
- 2318704 DCB MDAS 16 slice See LD for details Part #: 2318704 Part Type: GP CONSUMABLE - RETURNABLE Part Weight: -14549.85 LBR Repairable: No Returnable: Yes Hazardous Material: N Description: 16 SLICE DCB MDAS FRU: Yes Status: Orderable Parts Offered as: Current: 2318704 Harvested: Not Available Exchange: Not Available Comments: FLS - 16 SLICE DCB MDAS ADDED PER RSPL. USED ON H16. MSA8429 06/26/02
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required