GE Healthcare · CTLightSpeed244609
LightSpeed 4.X - GRE
Symptom
Invalid or mistached DAS type messages appear when tube warm up is attempted. Invalid or mistached DAS type messages appear when tube warm up is attempted
Error message
Error: 244609 WARNING - Invalid DAS is detected. The gantry can be down or the DAS can be bad See LD for complete error Mon Jan 23 07:02:57 2006 Error: 244609 Host: wct1 Process: fwmgr File: HwSetUp.c Line: 4117 Function: No system function reported Scan Type: Not scanning Scan: 0/0/0 Scan Seq Id: 0 WARNING - Invalid DAS is detected. The gantry can be down or the DAS can be bad
Applies to
- Models
- LightSpeed
- Product
- LightSpeed 4.X - GRE See Long Description for detailed information - H2_P_M16 Product Name: LightSpeed 4.X with GRE Product Photographs GE HEALTHCARE Method of remote or software identification - See attachment below for a complete list of software commands and typical results 1. check_config 2 . swhwinfo “swhwinfo” will have the following switches: · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) -h or –help = Outputs complete details about all the available switches associated with swhwinfo swhwinfo –c –m –o -da List of Supported Console types: GOC3 (GRE) GOC4 (GRE) List of software the product can run: 07MW11.10 SP2.0, SP3.0, SP3.1 FAQ's Table weight limit - 180 kg (400 lbs) normal operation - 205 kg (450 lbs) maximum allowed with decreased accuracy Tube type - D3186T D3186T(green) Current production - M4 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Component
- 5270886 Small Signal Brush Tip Kit and 5350798 high silver tip kit Replace 2349604 SIGNAL BRUSH TIP COLLECTOR 4 signal brush tips and caps per kit See Long Description for details Refer to attached service note for Important details SN09092411a This full set of Signal Brush Part Number has been divided up to two smaller Part numbers refer to SN09092411a 5270886 Signal brush kit quantity of 4 tips and caps (order 5 to get full set of 20 signal brush tips). 5350798 Red Type1 tip (High silver content) signal brush kit quantity of 4 tips and Caps _______________________________________________________________________________ Part #: 5270886 Part Type: GP CONSUMABLES Part Weight: 0.00 LBR Repairable: No Returnable: No Hazardous Material: N Description: Small Signal Brush Tip Kit FRU: Yes Status: Orderable Comments: 4 signal brush tips and caps per kit. Order a quantity of 5 for an entire set of signal brush tips. Kit does not contain type 1 red tip high silver content brush tips. Order part 5350798 for high silver tips. 212007413, Aug-2009 Per ECR 2087298, 212008898, 7/20/09. Part #: 5350798 Part Type: GP CONSUMABLES Part Weight: 0.00 LBR Repairable: No Returnable: No Hazardous Material: N Description: Small Signal Type 1 Brush Tip Kit FRU: Yes Status: Orderable Comments: 4 type 1 high silver content signal brush tips and caps per kit. 212007413, Aug-2009 Per ECR 2087298, 212008898, 7/20/09.
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required