Pular para o conteúdo principal
GE Healthcare · CTLightSpeed183127

Lightspeed 1.X - Octane

Symptom

Scanner Hardware Reset Failed. See LD Scanner Hardware Reset Failed Hardware Failed to reset message appeared on monitor AFTER boot-up WITH Error:178568 Failed to establish socket connection to STC in 10 seconds and errors 244553 and 244554 and power up self test fails A, B, and 9 on STC CPU board The following errors were in the gesyslog Thu Jan 13 04:18:32 2005 Error: 178568 Suite: RP25 Host: SBC0 Process: fwmgr File: NBSSocket.m 1.9 Method: connectToTarget Line: 315 Failed to establish socket connection to STC in 10 seconds Thu Jan 13 04:18:32 2005 Error: 244553 Suite: RP25 Host: SBC0 Process: fwmgr File: HwSetUp.m 1.18 Method: hardReset Line: 429 The STC controller has failed to reset. Thu Jan 13 04:18:32 2005 Error: 244554 Suite: RP25 Host: SBC0 Process: fwmgr File: HwSetUp.m 1.18 Method: hardReset Line: 442 The reset failure of the OBC may be due to the reset failure of the STC. Also saw failures A, B, and 9 on the power-up test on the STC CPU .

Error message

Error: 183127 FIL/MA board stuck bit See long description Tue Oct 28 13:12:38 2008 Error: 183127 Host: OBC Process: Control File: Line: 275 Function: XRAY GENERATION : MA Loop Scan: 0/0/0 Type: None/Unknown Exception Level: Pri/Soft Time: 615 Log Series: 2 FIL/MA board stuck bit error . Hardware address = 0xffdf47 Actual status = 0xba Expected status = 0xaa

Applies to

Models
LightSpeed
Product
Lightspeed 1.X - Octane 10.5_2.8.2I_H1.3M4 10.4.1_2.8.2I_H1.3M4 - obsolete Lightspeed 1.X - Octane Method of visual identification Method of remote or software identification - See attachment below for a complete list of software commands and typical results. check_config showprods|grep Apps note: if check_config does not work then the software is not up to date with all FMI's. Check FMI history and verify that all patches have been loaded since last software reload. Ref FMI 25269 List of Supported Console Types Octane with Pegasus Recon hardware. may have an 02 SBC - Run df -k to see if the sbc is partitioned. List of softwares the product can run 10.5_2.8.2I_H1.3M4 10.4.1_2.8.2I_H1.3M4 - obsolete FAQ's Table weight limit - Up to 400 lb. (180 kg) with ± .25 mm positional accuracy. 450Ib (205 kg) max weight Tube type - D3176T Current production details - See Technical Reference Manual found on the Common Document Library Tips and workarounds, where to find it - Common Document Library(CDL) . Search for the latest FMI software release. In the furnished materials will be the list of various manuals with part numbers for the various languages. Example: English versions 2211223-100 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
2154834 MA Board used on (CT/i,1.X-4.X) scanners See Long Description for more Part #: 2154834 Part Type: GP CONSUMABLE - RETURNABLE Part Weight: 1.81 LBR Repairable: No Returnable: Yes Hazardous Material: N Description: MA CONTROL BRD./CANCEL FRU: Yes Status: Orderable Comments: FLS - MA BOARD USED ON FAST RHAPSODE FMI#25238 UPGRADES THIS BD. TO A REV.1 ONE USE LIGHTSPEED ULTRA HQT WARNING!! MATERIAL MUST BE PURCHASED FROM MFGING DO NOT GO TO VENDOR AS IT MUST BE TESTED IN MFGING BEFORE SHIPPED TO FIELD VERY VERY HIGH DOA RATE. MS08992 GVR 7/31/00GELS
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required

Related GE Healthcare CT issues