Pular para o conteúdo principal
GE Healthcare · CTLightSpeed24518317954880005302724505036505

LightSpeed 4.X - GRE

Symptom

Scanner Hardware fault during Helical Scan data acquisition Problem is becoming less intermittent and exam sometimes proceeds normally once the resume button is selected. The problem also occurs during warm-up scans and fastcals

Error message

Error: 245183 Did not receive reply from a request. See long description for a sample error message. Tue Jan 5 09:18:44 2010 Error: 245183 Host: cbct001 Process: scanmgr File: firmware if.cxx Line: 1554 Function: System Control : Scan Control Scan Type: Helical Scan: 17954/3/2 Scan Seq Id: 88000 Did not receive reply from a request. Reply Event : 53027 Reply Source: scan control Error: Timed out while attempting to receive a message.

Error: 245050 Error verifying that Data Acquisition is ready for scanning Open Long Description to see error examples Error: 245050 Error verifying the Data Acquisition is ready for scanning. More than one error text exists for this error below: Fri Jul 1 16:51:14 2005 Error: 245050 Host: ct02 Process: scanmgr File: start_scanning.cxx Line: 231 Function: System Control : Scan Control Scan Type: Not scanning Scan: 483/1/0 Scan Seq Id: 1 Error -1 verifying that Data Acquisition is ready for scanning Wed Mar 3 16:56:16 2010 Error: 245050 Host: uhctlspro Process: scanmgr File: start_scanning.cxx Line: 250 Function: System Control : Scan Control Scan Type: Not scanning Scan: 36505/1/0 Scan Seq Id: 0 Error 200001821 verifying that Data Acquisition is ready 

Applies to

Models
LightSpeed
Product
LightSpeed 4.X - GRE See Long Description for detailed information - H2_P_M16 Product Name: LightSpeed 4.X with GRE Product Photographs GE HEALTHCARE Method of remote or software identification - See attachment below for a complete list of software commands and typical results 1. check_config 2 . swhwinfo “swhwinfo” will have the following switches: · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) -h or –help = Outputs complete details about all the available switches associated with swhwinfo swhwinfo –c –m –o -da List of Supported Console types: GOC3 (GRE) GOC4 (GRE) List of software the product can run: 07MW11.10 SP2.0, SP3.0, SP3.1 FAQ's Table weight limit - 180 kg (400 lbs) normal operation - 205 kg (450 lbs) maximum allowed with decreased accuracy Tube type - D3186T D3186T(green) Current production - M4 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required

Related GE Healthcare CT issues