GE Healthcare · MRSigna2225734
Stops in Middle of Scans Due to Low output of one of the SRFD amps.Scan Stops in Middle of Scans.Replace Analogic RF Amp…
Stops in Middle of Scans Due to Low output of one of the SRFD amps.Scan Stops in Middle of Scans.Replace Analogic RF Amplifier SRFD wqr.Signa LX.2225734 RF Interface (RFI) Mismatch Fault.
Symptom
Scan Stops in Middle of Scans See Long Description Hard failure during scanning, APS works ok, AP recon error watching APS window.
Error message
2225734 RF Interface (RFI) Mismatch Fault. The Dump Resistor in the RFI has overheated. 2225734 "RF Interface (RFI) Mismatch Fault. The Dump Resistor in the RFI has overheated. Possible causes: improper calibration, missing or improper cables between RFI and amplifier(s), defective RFI or defective Analogic RF Amplifier(s).
Applies to
- Models
- Signa
- Product
- Signa LX Signa LX Product can be used with 8.x and 9.x systems.
- Software
- 9.1 For Signa LX product only
- Component
- 2224562-2 1.5T RF AMP POWER MODULE W/ FRU PKG (SRFD)
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- Running grafidy3, and find there are spikes on the waveforms in the manual prescan display, both windows.
- Fails to Scan with TORSO Coil.
- The Carriage Fails to Advance to Latch to the Table and to Retract When it is Raised, Undocked or Lowered.
- AE title has changed on destination host.
- CWA ID100315. Auto and Manual Prescan Fails.
- What is the Acoustic Noise Level in the Ovation Magnet.