Pular para o conteúdo principal
GE Healthcare · MRSigna

Cold Head is Making Banging or Ratcheting Noises Due to Failed Cold Head.Cold Head is Making Banging or Ratcheting Noise…

Cold Head is Making Banging or Ratcheting Noises Due to Failed Cold Head.Cold Head is Making Banging or Ratcheting Noises.Second Opinion Part Number 2244334 SINGLE STAGE CONFIGURED COLD HEAD,wqr.Magne…

Symptom

Cold Head is Making Banging or Ratcheting Noises. Cold Head is Making Banging or Ratcheting Noises

Applies to

Models
Signa
Product
Magnet HFO. Software Version Concept should NOT be selected for Magnet Products Serial # (Txxx) Field strength: 0.7 T He Vessel Pressure >0.9 psig but <6.25 psig Shield Temp for Cryo-cooler coldhead < 40 K Temp for Shield-cooler Coldhead < 35 K Cryocooler 2nd stage Temp > 3.8K but < 4.5K Recondenser Temp > 4.2K but < 4.6K Delta Temp (Coldhead - Recondensor) < 0 but < 0.3K Min. LHe Fill Level 75% of total vessel Min. LHe Ramp Level 100% Helium Volume 340 Liters Burst Disk 2253802-2 Baffle Assy 2226764
Component
Second Opinion Part Number 2244334 SINGLE STAGE CONFIGURED COLD HEAD. Second Opinion Part Number 2244334 COLDHEAD, SINGLE STAGE CONFIGURED. Used exclusively on HFO (OpenSpeed) 0.7T Magnets. See comments in resolution concept for second opinion requirements before ordering this part. Second Opinion Part Number 2244334 COLDHEAD, SINGLE STAGE CONFIGURED. Used exclusively on HFO (OpenSpeed) 0.7T Magnets. See comments in resolution concept for second opinion requirements before ordering this part.
Area
MR

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.