GE Healthcare · MRSigna224859228727224857922485841207812070

TPS Reset Fails Due to Loose Data Cable Connection Underneath SCIM.

Symptom

Fails TPS Reset when booting up to Applications or when Resetting TPS See Long Description When the MR system is booting up to applications it will fail TPS reset but Applications will successfully boot. With Applications up and "Reset TPS" has been selected, TPS fails to download successfully.

Error message

2248592 SCIM Configuration Failed SCIM link errors caused the configuration to fail 2248592 Host: PMR5_OC0 Process: NSP File: SCP:OpWorkspaceAutoConfig Line: 364 Exception Class: Unknown Severity: Unknown Function: Peripherals: Communications PSD: Gradient Echo Coil: HEAD Scan: 28727/1 SCIM Configuration Failed. SCIM link errors caused the configuration to fail. Please fix and perform a TPS Reset

2248579 A timeout occurred when trying to receive a SCIM packet 2248579 Host: PMR5_OC0 Process: NSP File: SCP:KeyboardLinkResources Line: 471 Exception Class: Unknown Severity: Unknown Function: Peripherals: Communications PSD: Gradient Echo Coil: HEAD Scan: 28727/1 SCIM Communication Error. A timeout occurred when trying to receive a SCIM packet. Opcode = 0x60 Check cable connection between the SCIM and WIM and cable connection between the SCIM and System Cabinet.

2248584 SCIM Communication Error. Thu Apr 10 07:21:09 2008 2248584 2 NSP SCP:KeyboardLinkHWO 1920 Exception Class: Unknown Severity: Unknown Function: Peripherals: Communications SCIM Communication Error. The SCIM failed to respond to a watchdog packet from the SCP. Watchdog packets waiting for a response: 0x0000 Check cable c

Applies to

Models
Signa
Product
Signa HDxT This product is for 1.5T Signa HDxT systems with 15.0 software and HDi systems with 15.0 software. This product is for 1.5T Signa HDxT systems with 15.0 software and HDi systems with 15.0 software.
Software
15.0 For Product Signa HDxT and HDi only. -- NOT to be used with HDe. -- For Product Signa HDxT and HDi only. -- NOT to be used with HDe. --
Area
MR

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required