Pular para o conteúdo principal
GE Healthcare · CTLightSpeedVCT2601074641123627685

LightSpeed 7.x (VCT, VCT Select, Pro32)

Symptom

A grinding, rumbling, squealing, clicking, or other noise is heard only during gantry rotation - document in RFS - critical issue see long description for more detail. A grinding,rumbling, squealing, clicking, or other noise is heard only during gantry rotation. Gantry speed or velocity errors are NOT reported to the customer or the gesyslog. System will still scan ok but customer just complains of excessive noise from gantry. Sound volume or quality can change with rotation speed.

Error message

Error: 260107464 The Axial AC Drive reported fault code: F:64 Power Overload This fault code is undefined and may indicate a defective Drive. Retry operation. See Long Description for error example Error: 260107464 The Axial AC Drive reported fault code: F:64 Power Overload 1123627685 3 1 Tue Aug 9 22:48:05 2005 260107464 4 tgp tgp_subsys AX2fault.c 195 Tue Aug 9 22:48:05 2005 Host : Table Gantry Processor Ermes # : 260107464 Exception Class : Abort Severity : Primary File : AX2fault.c Line# : 195 Function : AXIAL CONTROL : Axial Communications Scan Type : Axial Scan: 745/2/31 %s The Axial AC Drive reported fault code: F:64 Power Overload This fault code is undefined and may indicate a defective Drive. Retry operation. If problems persist contact the factory for instructions.

Applies to

Models
LightSpeed, VCT
Product
LightSpeed 7.x (VCT, VCT Select, Pro32) V40_H_V64_G_GTM V40_H_V64_G_GTL V40_H_V32_G_GTM V40_H_V32_G_GTL V40_H_H64_G_GTM V40_H_H64_G_GTL V40_H_HS64_G_G Product Name: LightSpeed 7.x (VCT, VCT Select, Pro32) Product photographs Method of remote or software identification - See attachment for a complete list of software commands and typical results. check_config swhwinfo “swhwinfo” will have the following new switches available with the VCT M4 and PF 1.5 M3 software release (software release deploying this “swhwinfo” capability is “06MW03.?”): · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) · -h or –help = Outputs complete details about all the available switches associated with swhwinfo · swhwinfo –c –m –o -da List of Supported Console Types GOC5 Will have a VDARC and VIG GOC6 Will have a VDIG List of softwares the product can run 09MW08.10 09MW08.11 08MW33.2 07MW18.4 06MW03.4 05MW31.6 FAQ's Table weight limit - Up to 500 lb. (227 kg) with ± .25 mm positional accuracy Tube type - D3194T D3193T(green) Current production - M4 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system. Done acording to template
Component
2277225 Axial Brake assembly ( Holding Brake for Axial Drive ) See Long Description for more info 2277225 Axial Brake assembly ( Holding Brake for Axial Drive ) Part #: 2277225 Part Type: GP CONSUMABLES Part Weight: 2.27 LBR Repairable: No Returnable: No Hazardous Material: N Description: BRAKE ASSY, H2 AX DRIVE MOTOR FRU: Yes Status: Orderable Comments: FLS - H2 GANTRY: MUST ORDER FROM HOLT ELECTRONIC. MS26561 DSM 1/31/
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required