GE Healthcare · CTLightSpeed18480235462

LightSpeed 2.X (Plus, QX/i) - Octane

Symptom

CONSOLE LOOSES COMUNICATION WITH GANTRY

Error message

Error: 184802 Communication Facility Error: 2103 Dest uuid: 0 Msg Code: 7021 See LD for complete error Fri Mar 10 10:50:32 2006 Error: 184802 Host: ETC Process: Table Controls File: tc_exec.c 1.27 1/15/96 Line: 219 Function: No System Function Reported Scan: 35462/1/0 Type: None/Unknown Exception Level: Pri/Most Severe Time: 10:50:32:880 Log Series: 267 Communication Facility Error: 2103 Dest uuid: 0 Msg Code: 7021

Applies to

Models
LightSpeed
Product
LightSpeed 2.X (Plus, QX/i) - Octane Always S-DAS 2.4.2_H2.4M5 LightSpeed 2.X (Plus, QX/i) - Octane (Pegasus or Rhapsode IG) - Always S-DAS Method of visual identification Method of remote or software identification - See attachment below for a complete list of software commands and typical results. check_config List of Supported Console Types Octane with Pegasus Recon hardware. List of softwares the product can run 2.4.2_H2.4M5 FAQ's Table weight limit - Up to 400 lb. (180 kg) with ± .25 mm positional accuracy. 450Ib (205 kg) max weight Tube type - D3186T Current production details - See Technical Reference Manual found on the Common Document Library Tips and workarounds, where to find it - Common Document Library(CDL) . Search for the latest FMI software release. In the furnished materials will be the list of various manuals with part numbers for the various languages. Example: English versions 2211223-100 FEATURES/OPTIONS STD/OPTION Cat # Digital Laser Camera I/F B7700PD Smart Helical (s/w only) Standard B7530TK Fast Recon (s/w only) Option B7700JF BMD (requires B7710MP or MN) Option B7710MK Smart Prep Option B7530SP Smart Score ? ? 3D Analysis (Requires D3D) Option B7540TD Navigator (s/w only) Option B7540N DentaScan (s/w only) Option B7540SD Direct 3D (Requires B7710SA) Option B7810SS HIS/RIS Option (h/w and s/w) Option B7500PN Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
5139480-2 8-port ethernet switch See long description Part #: 5139480-2 Part Type: GP CONSUMABLES Part Weight: 0.00 LBR Repairable: No Returnable: No Hazardous Material: N Description: 8-Port Switch Kit with 4 cushions FRU: Yes Status: Orderable Comments: Added per ECO# 2038731 HDe, 24-Jan-2007 305000160/
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.