Pular para o conteúdo principal
GE Healthcare · CTBrightSpeedHiSpeed2601182155036826010011710697

BrightSpeed HPower, Delight, Velocity or MERI HPower

Symptom

Scanner Hardware Stopped Scan with High Voltage errors

Error message

Error: 260118215 (60-0318) kV did not reach 75 percent after 20ms See Long Description for full error Sat Mar 25 15:35:28 2006 Error: 260118215 Host: orp Process: orp_subsys File: gen_exec.c Line: 2150 Sat Mar 25 15:35:27 2006 Host : Jedi Ermes # : 260118215 Exception Class : Abort Severity : Primary File : gen_exec.c Line# : 2150 Function : No System Function Reported Scan Type : Static Scan: 50368/1/3 (60-0318)kV did not reach 75 percent after 20ms. Indicates that there was no HV ON signal generated for exposure time count-up.

Error: 260100117 Timeout on Exposure Complete Data containing the Generator Xray Scan Information (Avg mA). See long description for sample error. Tue Jun 22 09:46:30 2010 Error: 260100117 Host: tgp Process: tgp subsys File: scTrigSeqActions.c Line: 1737 Exception Class : Abort Severity : Secondary Function : XRAY GENERATION : Final Exposure Command Scan Type : Axial Scan: 10697/2/4 Timeout on Exposure Complete Data containing the Generator Xray Scan Information (Avg mA). HW Error: Possible hardware problem. Timeout is: 650 mS.

Applies to

Models
BrightSpeed, HiSpeed
Product
BrightSpeed HPower, Delight, Velocity or MERI HPower HP_P_G4_G_HPT Excel HP_P_G8_G_HPT Edge HP_P_G16_G_HPT Elite HP_P_G16_G_GMV Elite HP_P_P16_G_GMV Elite HP_P_M16_G_HPT BrightSpeed Method of remote software hardware identification check_config swhwinfo List of Supported Console Types : HP_P_G16_ G _HPT Note: G could be GRE, AIO or TIO - check with: cat /usr/g/config/INFO | grep CONSOLETYPE GOC4 AIO TIO List of softwares the product can run: 08BW20.1 07MW11.10 08BW44.1 ( Delight ) 09HW30.4 ( Velocity ) 09BW35.11 ( MERI - Pancake DAS ) FAQ's 1. All BrightSpeed Power Systems use the Performix tube D3186T HP_P_G4_G_HPT (Excel - H-Power Table) HP_P_G8_G_HPT (Edge -H-Power Table ) HP_P_G16_G_HPT (Elite -H-Power Table ) HP_P_G16_G_GMV (Elite -Global Table Mid Value - modified GT1700 no IMS ) Note: G could be GRE, AIO or TIO (check with: cat /usr/g/config/INFO | grep CONSOLETYPE ) BrightSpeed Elite Power Systems will only be sold as Velocity System starting in 2Q/2010. Delight Elite Power being phased out 1Q/2010 BrightSpeed Power 4 & 8 slice systems have been discontinued due to restructuring of the product lineup. Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
2401826-2 Jedi 60 Shielded Power Unit / High Voltage Tank Assembly FRU Package for G-DAS systems Includes: 2377668-2 Jedi 60 Power Unit 2186730 Jedi 60 HV Tank Shielding improved for G-DAS systems See LD for details 2401826-2 JEDI 60DC SHIELDED POWER UNIT / HV TANK HLB ASSY With FRU PACKING FOR G-DAS SYSTEMS (Backward compatible with M-DAS systems) Part #: 2401826-2 Part Type: GP CONSUMABLES Part Weight: 2.43 LBR Repairable: No Returnable: No Hazardous Material: N Description: JEDI 60DC SHIELDED POWER UNIT / HV TANK HLB ASSY With FRU PACKING FRU: Yes Status: Orderable Comments: Revised per to MFG confirmation. This is part of Jedi Collector # 2371333-2 INCLUDES: 2377668-2 JEDI 60DC HLB POWER UNIT with INVERTER and PPC KV Control 2186730 JEDI 60 STD HV TANK
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.