Pular para o conteúdo principal
GE Healthcare · CTLightSpeed260005502

LightSpeed 3.X (Ultra, Plus, QX/i) - GOC2

Symptom

Replaced Pegasus board and still cannot cu ice

Error message

Error: 260005502 failed to respond to procICE: status = -1 See long description for full error Sat May 24 09:46:50 2008 Error: 260005502 Host: ice-post Process: vxmon_post File: vxMonitorMain.c Line: 761 failed to respond to procICE: status = -1

Applies to

Models
LightSpeed
Product
LightSpeed 3.X (Ultra, Plus, QX/i) - GOC2 This is HP CPU host running LINUX with Pegasus IG. 308L.8_H3.3M4 ~ 4 or 8 slice MDAS. ~ Could be an Ultra, Plus or QXi system. Must run swokinstall -p to verify. ~ See Long description for details . LightSpeed 3.X MDAS4 Gantry and GOC2 Gantry - 2281177-2, DAS - 2272768, Console – 2341104-2, Table - 2276776, PDU - 2269902 Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
2333829-2 SCSI CABLE KIT, GOC-LINUX See long description Part #: 2333829-2 Part Type: GP CONSUMABLES Part Weight: 2.18 LBR Repairable: No Returnable: No Hazardous Material: N Description: SCSI CABLE KIT, GOC-LINUX FRU: Yes Status: Orderable Comments: FLS - SCSI CABLE KIT, GOC-LINUX. AILTO P/N 2315653. MS20544 ON 3/21/03.