MRSPT Head System Gain Fails with Low Signal Due To reversed cable on 16 channel switch.

Symptom

MRSPT Head System Gain Fails with Low Signal

Error message

2224580 Auto Prescan failed. Signal was too small to process - use Manual Prescan 2224580 Host: hopgemr1a Process: NSP File: APS:aps_cf Line: 11 Auto Prescan failed. Signal was too small to process - use Manual Prescan

2224725 Low signal detected. Please check the signal level for receiver X (or receiver X while using Manual Prescan) See Long Description Error: 2224725 Host: hopgemr1a Process: NSP File: APS:aps_frame_range_check Line: 10 Low signal detected. Please check the signal level for receiver X (or receiver X while using Manual Prescan).

Applies to

Models
Signa
Product
3.0T Signa HDxT For 3.0T Signa HDxT systems with 15.0 software only. For 3.0T Signa HDxT systems with 15.0 software only.
Software
15.0 For Product Signa HDxT and HDi only. -- NOT to be used with HDe. -- For Product Signa HDxT and HDi only. -- NOT to be used with HDe. --
Area
MR