Pular para o conteúdo principal
GE Healthcare · CTLightSpeed200520009200501502200108060123621136192391551637

LightSpeed 4.X - GRE

Symptom

Software shut itself down after clicking confirm to start scan series Software shut itself down after clicking confirm to start helical scan series

Error message

Error: 200520009 Process: environmentMgr EnvironmentMgr received SIGPIPE, broken pipe Thu Jan 13 13:04:07 2011 Error: 200520009 Host: ct03 Process: environmentMgr File: fsaMgr.c Line: 227 EnvironmentMgr received SIGPIPE, broken pipe

Error: 200501502 UifXIoErrorHandler: Encountered an X library I/O error. errno = xx Process will intentionally dump core. See Long Description for a complete error. Tue Nov 8 12:03:08 2005 Error: 200501502 Host: ct03 Process: environmentMgr File: UifXErrorHandlers.c Line: 124 Function: No system function reported UifXIoErrorHandler: Encountered an X library I/O error. errno = 32 Process will intentionally dump core.

Error: 200108060 Localizer invalid for graphicRx See LD for complete error Mon Jun 5 21:59:52 2006 Error: 200108060 Host: hmract2 Process: scanRx File: UIRx.cxx Line: 12362 Function: Data Acquisition : OC Processing Localizer invalid for graphicRx. Localizer azimuth = 90 Localizer landmarkuid = 1.2.840.113619.2.55.3.2391551637.5625.1149512358.895.82 64.0.11 Prospective landmarkuid = 1.2..840.113619.2.55.3.2391551637.5625.1149512358.895. 0.0.11

Error: 200002352 A HOST_SHUTDOWN was initiatied by the death of msgView See Long Description for

Applies to

Models
LightSpeed
Product
LightSpeed 4.X - GRE See Long Description for detailed information - H2_P_M16 Product Name: LightSpeed 4.X with GRE Product Photographs GE HEALTHCARE Method of remote or software identification - See attachment below for a complete list of software commands and typical results 1. check_config 2 . swhwinfo “swhwinfo” will have the following switches: · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) -h or –help = Outputs complete details about all the available switches associated with swhwinfo swhwinfo –c –m –o -da List of Supported Console types: GOC3 (GRE) GOC4 (GRE) List of software the product can run: 07MW11.10 SP2.0, SP3.0, SP3.1 FAQ's Table weight limit - 180 kg (400 lbs) normal operation - 205 kg (450 lbs) maximum allowed with decreased accuracy Tube type - D3186T D3186T(green) Current production - M4 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required

Related GE Healthcare CT issues