Pular para o conteúdo principal
GE Healthcare · CTLightSpeedVCT260003100

LightSpeed 3.X (Ultra, Plus, QX/i) - GRE

Symptom

"Scan hardware Stopped Scan" with collimator errors. "Scan hardware Stopped Scan" with collimator errors

Error message

Error: 260003100 CAM B Motion With Out Command Exceeded The Error Tolerance. Desired Position : 251um 797cnt Current Position : 415um 763cnt Error Tolerance: 44um Mon Dec 6 13:01:26 2004 Error: 260003100 Host: HSC Process: CCB Firmware File: ccb_drive_timers.c Line: 277 Running diagnostic: Collimator Functional Function: XRAY GENERATION : Xray Collimation and Filtration Scan: 0/0/0 Type: None/Unknown Exception Level: Sec/Soft Time: 13:01:27:185 Log Series: 42 CAM B Motion With Out Command Exceeded The Error Tolerance. Desired Position : 251um 797cnt Current Position : 415um 763cnt Error Tolerance: 44um

Applies to

Models
LightSpeed, VCT
Product
LightSpeed 3.X (Ultra, Plus, QX/i) - GRE Always 4 or 8 slice MDAS. H2_P_M4 H2_P_M8 LightSpeed 3.X - GRE Method of visual identification Small rounded gantry similar in appearance to LS 2.X and 4.X. Always has an MDAS4 or MDAS8 Method of remote or software identification - See attachment below for a complete list of software commands and typical results. check_config “swhwinfo” will have the following new switches available with the VCT M4 and PF 1.5 M3 software release (software release deploying this “swhwinfo” capability is “06MW03.?”): · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) -h or –help = Outputs complete details about all the available switches associated with swhwinfo swhwinfo –c –m –o -da Always has an MDAS4 or MDAS8 List of Supported Console Types GOC3 Will have a DARC, SDDA and IG GOC4 Will have a DARC 2 and IG List of softwares the product can run 07MW11.10 SP3 06MW29.7 - obsolete 06MW03.5 - obsolete 05MW14.5 - obsolete 04MW44.9 - obsolete 04MW13.7 - obsolete FAQ's Table weight limit - Up to 400 lb. (180 kg) with ± .25 mm positional accuracy. 450Ib (205 kg) max weight Tube type - D3186T Current production details - See Technical Reference Manual found on the Common Document Library Tips and workarounds, where to find it - Common Document Library(CDL) . Search for the latest FMI software release. In the furnished materials will be the list of various manuals with part numbers for the various languages. Example: English versions 2211223-100 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
5140001-3 (replaces 5140001-2) Helios 3 collimator with notched filler bar and rating plate (used on 1.X-4.X) See Long Description for more info Part #: 5140001-3 Part Type: GP REPAIRABLES Part Weight: 115.00 LBR Repairable: Yes Returnable: Yes Hazardous Material: N Description: helios 3 collimator with notched filler bar and rating plate FRU: Yes Status: Orderable Comments: ** FE Assist ** 1.x 2.x 3.x 4.x started with Ultras added minor quality improvements to the dash 2 and is backwards compatible (12/6/06 212001261) Part #: 5140001-2 Part Type: GP REPAIRABLES Part Weight: 0.00 LBR Repairable: Yes Returnable: Yes Hazardous Material: N Description: helios 2 collimator with notched filler bar and rating plate FRU: Yes Status: Orderable Comments: This part is replacement for 2214768-2, please do not use stock of 2214768-2. ms13452 050516 Can use the 5140001-3 if no Stock of the -2 (9/26/05 212001261)
Area
CT

Full cause & step-by-step resolution