CWA Auto Prescan Failure Due to metal present in anatomy, bandages, braces, dental work, etc.

Symptom

Unable to scan with Power Monitor Faults

Error message

2248816 RF Interface (RFI) High Voltage Relay Fault 2248816, RF Interface (RFI) High Voltage Relay Fault. Possible causes: RF Monitor Cable to System Support Module (SSM) is disconnected or defective.

2248815 Power Monitor 2 Fault. RF Transmit power has reached safety limit allowed for scan parameters and patient weight. Check patient weight entered. This could also occur when power cycling SSM.

2248814 Power Monitor 1 Fault. RF Transmit power has reached safety limit allowed for scan parameters and patient weight. Check patient weight entered. This could also occur when power cycling SSM.

Applies to

Product
Open Speed with EXCITE For HFO4 and HFO5 only
Software
OPENMR4-HFO4 Software version for HFO4 system, which is OpenSpeed with Excite Technology
Area
MR