Symptom

Scanner Hardware Stopped Scan. Scanner Hardware Stopped Scan

Error message

Error: 182126 LAN Watchdog TIMEOUT with OBC (uuid 193) See LD for complete error Wed Sep 20 08:39:07 2006 Error: 182126 Host: STC Process: Master Exc Handler File: exm_lan_wdog.c Line: 851 Function: No System Function Reported Scan: 50091/8/0 Type: None/Unknown Exception Level: Secondary Time: 08:39:01:815 Log Series: 724 LAN Watchdog TIMEOUT with OBC (uuid 193) (SW opening abort relay). HW/SW Error: Did not receive LAN Watchdog message from OBC. (Last Wdog Probe Sent:491163 Last Response Rcvd:490966 (in Ticks)).

Applies to

Models
LightSpeed
Product
LightSpeed 2.X (Plus, QX/i) - Octane Always S-DAS 2.4.2_H2.4M5 LightSpeed 2.X (Plus, QX/i) - Octane (Pegasus or Rhapsode IG) - Always S-DAS Method of visual identification Method of remote or software identification - See attachment below for a complete list of software commands and typical results. check_config List of Supported Console Types Octane with Pegasus Recon hardware. List of softwares the product can run 2.4.2_H2.4M5 FAQ's Table weight limit - Up to 400 lb. (180 kg) with ± .25 mm positional accuracy. 450Ib (205 kg) max weight Tube type - D3186T Current production details - See Technical Reference Manual found on the Common Document Library Tips and workarounds, where to find it - Common Document Library(CDL) . Search for the latest FMI software release. In the furnished materials will be the list of various manuals with part numbers for the various languages. Example: English versions 2211223-100 FEATURES/OPTIONS STD/OPTION Cat # Digital Laser Camera I/F B7700PD Smart Helical (s/w only) Standard B7530TK Fast Recon (s/w only) Option B7700JF BMD (requires B7710MP or MN) Option B7710MK Smart Prep Option B7530SP Smart Score ? ? 3D Analysis (Requires D3D) Option B7540TD Navigator (s/w only) Option B7540N DentaScan (s/w only) Option B7540SD Direct 3D (Requires B7710SA) Option B7810SS HIS/RIS Option (h/w and s/w) Option B7500PN Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
2224296-3 Artysen board for 1X/2X/3X/4X used in the ETC/STC/OBC See Long Description for FEMC information and board locations Part #: 2224296-3 Part Type: GP REPAIRABLES Part Weight: 1.09 LBR Repairable: Yes Returnable: Yes Hazardous Material: N Description: VE-GE, 8MB, Artesyn CPU Bd FRU: Yes Status: Orderable Parts Offered as: Current : 2224296-3 Harvested : Not Available Exchange : Not Available Comments: ** FE ASSIST ** Repair operations has found that 68% of the time the Artesyn board (2224296-3) is returned for repair, it has No Defect Found. Check switch settings (212001261 2/12/09)
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required

Related GE Healthcare CT issues