Symptom

Customer was not able to remote tilt gantry. In many cases , gantry will tilt from the Gantry Controls.

Error message

Error: 182886 'The Remote Tilt Input to the TNC is STUCK' See LD for full detail Error 182886 'The Remote Tilt Input to the TNC is STUCK' Wed Jun 7 04:42:36 2006 Error: 182886 Host: ETC Process: Gantry Display File: gd tnc erlog.c Line: 639 Function: No System Function Reported Scan: 0/0/0 Type: None/Unknown Exception Level: Pri/Soft Time: 04:42:31:470 Log Series:7 The Remote Tilt Input to the TNC is STUCK.

Error: 182826 Communication Error: Communication Packet with the GANTRY DISPLAY. Open Long Description for full error Wed Dec 22 20:49:26 2004 Error: 182826 Host: ETC Process: Gantry Display File: gd_tnc_driver.c Line: 477 Function: No System Function Reported Scan: 0/0/0 Type: None/Unknown Exception Level: Pri/Soft Time: 20:48:40:485 Log Series: 31 Communication Error: Communication Packet with the GANTRY DISPLAY. Message Code: 2. (Data = 0x20, 0x0, 0x0, 0x0).

Error: 182825 Communication Problems with the GANTRY CONTROL See Long Description for full error Tue Dec 1 17:17:45 2009 Error: 182825 Host: ETC Process: Gantry Display File: gd_tnc_driver.c Line: 471 Function: No System Function Reported Scan: 0/0/0 Type: None/Unknown Exception Level: Pri/Soft Time: 17:17:39:290 Log Se

Applies to

Models
LightSpeed, VCT
Product
LightSpeed 4.X - GRE See Long Description for detailed information - H2_P_M16 Product Name: LightSpeed 4.X with GRE Product Photographs GE HEALTHCARE Method of remote or software identification - See attachment below for a complete list of software commands and typical results 1. check_config 2 . swhwinfo “swhwinfo” will have the following switches: · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) -h or –help = Outputs complete details about all the available switches associated with swhwinfo swhwinfo –c –m –o -da List of Supported Console types: GOC3 (GRE) GOC4 (GRE) List of software the product can run: 07MW11.10 SP2.0, SP3.0, SP3.1 FAQ's Table weight limit - 180 kg (400 lbs) normal operation - 205 kg (450 lbs) maximum allowed with decreased accuracy Tube type - D3186T D3186T(green) Current production - M4 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Area
CT

Full cause & step-by-step resolution