GE Healthcare · CTLightSpeedVCT
LightSpeed 7.x (VCT, VCT Select, Pro32)
Symptom
All Lights on the Gantry Display Lit After Switch Off the OC Power See Long Description for Details... All Lights on the Gantry Display Lit After Switch Off the OC Power . When console On/Off swith is turned off, all of the lights on the gantry display lite up
Applies to
- Models
- LightSpeed, VCT
- Product
- LightSpeed 7.x (VCT, VCT Select, Pro32) V40_H_V64_G_GTM V40_H_V64_G_GTL V40_H_V32_G_GTM V40_H_V32_G_GTL V40_H_H64_G_GTM V40_H_H64_G_GTL V40_H_HS64_G_G Product Name: LightSpeed 7.x (VCT, VCT Select, Pro32) Product photographs Method of remote or software identification - See attachment for a complete list of software commands and typical results. check_config swhwinfo “swhwinfo” will have the following new switches available with the VCT M4 and PF 1.5 M3 software release (software release deploying this “swhwinfo” capability is “06MW03.?”): · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) · -h or –help = Outputs complete details about all the available switches associated with swhwinfo · swhwinfo –c –m –o -da List of Supported Console Types GOC5 Will have a VDARC and VIG GOC6 Will have a VDIG List of softwares the product can run 09MW08.10 09MW08.11 08MW33.2 07MW18.4 06MW03.4 05MW31.6 FAQ's Table weight limit - Up to 500 lb. (227 kg) with ± .25 mm positional accuracy Tube type - D3194T D3193T(green) Current production - M4 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system. Done acording to template
- Component
- 2404320-2 CONSOLE INTERCOM PRESCRIBED-TILT HUB See Long Description for More ... 2404320-2 CONSOLE INTERCOM PRESCRIBED-TILT HUB Part #: 2404320-2 Part Type: GP CONSUMABLES Part Weight: 0.00 KGM Repairable: No Returnable: No Hazardous Material: N Description: CONSOLE INTERCOM PRESCRIBED-TILT HUB FRU: Yes Status: Orderable Comments: Per ECR 2062448, 212008898, 10/8/08. ** FE ASSIST ** CT/PET Discovery System's must use the P/N 2404320. CT System's must have FMI 25407 or FMI22918 for RT16/Xtra performed prior to installing 2404320-2. Verify Software Configuration before ordering/installing this FRU. To do this, open a Unix Shell and verify one of the following base software versions is installed: ctuser@hostname} swhwinfo -v 07MW11.10_SP2-0-26.Hardware info / 07MW18.4_SP2-0-26.Hardware info / 07BW08.4_SP2-0-26.Hardware info / 08BW17.7 Hardware info / 08BW20.x Hardware info (212001261 10/21/08)
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required