GE Healthcare · MRSigna2238645
Unable to Scan With Gram Error due to GRAM Module.
Symptom
Unable to Scan With Gram Error
Error message
2238645 The Z axis GRAM is reporting an under voltage fault. Thu Aug 14 09:54:12 1997 Suite: MRS1 Host: OC0 Proc: HPC_PORT_ID Error: 2238645 File: PMH.m 1.29 Method: gap Line: 1045 The Z axis GRAM is reporting an under voltage fault. Most likely cause is PSD prescription with demanding parameters or GRAM Hardware Fault.
Applies to
- Models
- Signa
- Product
- Signa LX Signa LX Product can be used with 8.x and 9.x systems.
- Software
- 9.1 For Signa LX product only
- Component
- 46-320147G2 GRAM Module GRAM Module 46-320147G2
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- Running grafidy3, and find there are spikes on the waveforms in the manual prescan display, both windows.
- Fails to Scan with TORSO Coil.
- The Carriage Fails to Advance to Latch to the Table and to Retract When it is Raised, Undocked or Lowered.
- AE title has changed on destination host.
- CWA ID100315. Auto and Manual Prescan Fails.
- What is the Acoustic Noise Level in the Ovation Magnet.