GE Healthcare · CTLightSpeed2600074215000026000738650056260007422260007353
LightSpeed 5.X (80/100kW)
Symptom
Scanner Hardware Stopped Scan with DAS parity errors
Error message
Error: 260007421 The DCB has an Outstanding SelfTest Failure which does NOT allow Patient Scanning. See Long Description for full error Wed Mar 2 16:57:53 2005 Error: 260007421 Host: orp Process: orp_subsys File: scanrx.c Line: 248 Wed Mar 2 16:57:51 2005 Host : Das Ermes # : 260007421 Exception Class : Soft Severity : Pri/Soft File : scanrx.c Line# : 248 Function : No System Function Reported Scan Type : Axial Scan: 50000/1/1 The DCB has an Outstanding SelfTest Failure which does NOT allow Patient Scanning. (The DCB current State: Service State ). Error: 260007386 DAS Has Detected An Error. Data Parity Error detected in the xx Data Stream(s) on Converter Board xx or the DCB Board. See LD for sample errors Wed Mar 23 15:46:11 2005 Error: 260007386 Host: DAS Process: DCB Firmware File: parity.c Line: 219 Function: DATA ACQUISITION : Scan Data Acquisition Scan: 214/5/1 Type: Helical Exception Level: Sec/Soft Time: 15:46:11:545 Log Series: 660 SCAN ABORT -- DAS Has Detected An Error. Data Parity Error detected in the 1B Data Stream (s) on Converter Board 33 or on Converter Board 41 . NOTE: Before troubleshooting, go through gesys.log (viewlog) and search for all “Data Parity Error” r
Applies to
- Models
- LightSpeed
- Product
- LightSpeed 5.X (80/100kW) LightSpeed 5.X (80/100kW) ~ Use with both MDAS and GDAS ~ GRE Console HP_H_G16 HP_H_M16 HP_O_M16 LightSpeed 5.X (100kW) GDAS16 with GRE Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Component
- 2318704 DCB MDAS 16 slice See LD for details Part #: 2318704 Part Type: GP CONSUMABLE - RETURNABLE Part Weight: -14549.85 LBR Repairable: No Returnable: Yes Hazardous Material: N Description: 16 SLICE DCB MDAS FRU: Yes Status: Orderable Parts Offered as: Current: 2318704 Harvested: Not Available Exchange: Not Available Comments: FLS - 16 SLICE DCB MDAS ADDED PER RSPL. USED ON H16. MSA8429 06/26/02
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required