GE Healthcare · MRSigna222381722196282227964222384822238472232902

Unable to reset TPS with AP Initialization errors Due to AP Board.

Symptom

Fails TPS Reset when booting up to Applications or when Resetting TPS See Long Description When the MR system is booting up to applications it will fail TPS reset but Applications will successfully boot. With Applications up and "Reset TPS" has been selected, TPS fails to download successfully.

Error message

2223817 TPS is Not Responding. Please Reset TPS Wed Apr 28 16:58:01 2004 Error: 2223817 Host: GEMS Process: NSP File: TardisReset.cc Line: 398 TPS is not responding. Please reset TPS.

2219628 The BAM Check Packet failed to reach the TPS Tue Jun 15 08:58:55 2004 Error: 2219628 Host: IRIS Process: scn File: SrxIPScreen.m Line: 1956 The BAM Check Packet failed to reach the TPS.

2227964 AP initialization MCOS error 0xff00010c Sat Jun 11 15:55:30 2005 Error: 2227964 Host: BWS2_OC0 Process: NSP e: TPS:ap_init_medcam Line: 499 AP initialization MCOS error 0xff00010c CE_MAP_ERROR: Error occurred mapping another Check that MEDCAM is present and jumpered for VME Address 0x40000000

2223848 TPS boot status packet has non-zero value. ermes is -1. Host: lx-bay14b Proc: NSP Error: 2223848 File: TardisReset.m Line: 1176 TPS boot status packet has non-zero value. ermes is -1.

2223847 TPS boot status packet is not received by NSP See Long Description Tue Feb 9 11:47:44 1999 Error: 2223847 Host: MRS2 Process: NSP File: TardisReset.m Line: 1177 TPS boot status packet is not received by NSP.

2232902 Processor Exception Frame, Bus Error: vector 2, interrupt level 0 Mon Nov 3 13:11:30 1997 Error: 22

Applies to

Models
Signa
Product
Signa LX Signa LX Product can be used with 8.x and 9.x systems.
Software
9.1 For Signa LX product only
Component
SecondOpinion Dual PPC Module 2222532-3 See comments in resolution concept for second opinion requirements before ordering this part See comments in resolution concept for second opinion requirements before ordering this part
Area
MR

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.