Pular para o conteúdo principal
GE Healthcare · CTDiscoveryVCT24528350003245061134200026010002850002

Scanner Hardware Stopped Scan.

Symptom

Scanner Hardware Stopped Scan. Scanner Hardware Stopped Scan

Error message

Error: 245283 Timeout (6000 ms) waiting on scan disk rezero response from dataacq See long description for full error Mon Nov 26 07:29:05 2007 Error: 245283 Host: svmcct16 Process: scanmgr File: scanmgr events.cxx Line: 1151 Function: System Control : Scan Control Scan Type: Not scanning Scan: 50003/2/0 Scan Seq Id: 0 Timeout (6000 ms) waiting on scan disk rezero response from dataacq.

Error: 245061 Process: scanmgr SCNMGR_ACQUISITION_DONE event was not received See LD for complete error Wed Feb 28 17:50:15 2007 Error: 245061 Host: ct99 Process: scanmgr File: norm_and_gen_timeouts.cxx Line: 90 Function: System Control : Scan Control Scan Type: Helical Scan: 0/0/0 Scan Seq Id: 1342000 SCNMGR_ACQUISITION_DONE event was not received

Error: 260100028 Scan Control timed out waiting for the Ready For Views message from the OC. See Long Description for complete error. Thu Apr 13 22:25:49 2006 Error: 260100028 Host: tgp Process: tgp_subsys File: scTrigSeqActions.c Line: 1616 Thu Apr 13 22:25:50 2006 Host : Table Gantry Processor Ermes # : 260100028 Exception Class : Abort Severity : Primary File : scTrigSeqActions.c Line# : 1616 Function : DATA ACQUISITION : Scan Data Acquisition Scan Ty

Applies to

Models
Discovery, VCT
Product
Discovery CT750 HD V40_PS_HD64_G_GTL V40_PS_HD64_G_GTM Discovery HD CT750 Method of visual identification HD gantry: Gray top and Gray Side covers with HD Markings Console Type: GOC6 has black SCIM and tower face Method of remote or software identification - See attachment below for a complete list of software commands and typical results. check_config swhwinfo “swhwinfo” will have the following new switches available · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) -h or –help = Outputs complete details about all the available switches associated with swhwinfo swhwinfo –c –m –o -da List of softwares the product can run (Consult attached FCT Software Cross Reference): 08MW39.4 - M3 (see attachement for M3 available options) 09MW29.14 - M4 (HD Options = M3 plus Gemstone Spectral Imaging, Gemstone Spectral Review and volume Helical Shuttle ) Example: Of difference with options software [insite@ct01 ~]$ swokinstall -p Product Software Option Keys: Gemstone Spectral Imaging Permanent HelicalShuttle Permanent Xtream Recon Permanent FAQ's Table weight limit: 500 lbs Tube Type - D3702T Green D3701T Current production - M4 Refer to product data sheets or Technical reference Manual, English 5317222-1EN Revision : 2 or later Refer to Application Tips and Work-Arounds Direction 5311815-1EN Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
5308760 Seagate - PN ST32050310NS - Baracuda ES SATA Family Hard Drives with Carrier Compatible - used On GOC6 Disk Array See Long description: Part #: 5308760 Seagate - PN ST32050310NS - Baracuda ES SATA Family Hard Drives with Carrier Compatible used on GOC6 Disk array ( 5308759 ). Part #: 5308760 Part Type: GP CONSUMABLES Part Weight: 0.00 LBR Repairable: No Returnable: No Hazardous Material: N Description: Seagate - PN ST32050310NS - Baracuda ES SATA Family Hard Drives with Carrier Compatible with 5308759 FRU: Yes Status: Orderable Comments: Per HDVCT RSPL, 212008898, 7/15/08. ** FE ASSIST ** This is a single Hard Drive used in the High Speed Disk Array on the GOC6 Console CT 750 HD and VCT (212001261 23-Jan-2009) ST32050310NS = 250 GB SATA II Hard Disk Drive
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required