Pular para o conteúdo principal
GE Healthcare · CTLightSpeed26010013810737607042600015121095852103260100137

LightSpeed 5.X (80/100kW)

Symptom

Gantry shutdown itself for 1 minute and then reset on its own. GANTRY SHUTDOWN ITSELF FOR 1 MINUTE AND THEN RESET ON ITS OWN BETWEEN SCOUTS AND AXIAL SCANS.

Error message

Error: 260100138 Firmware SHUTDOWN occurred caused by this Ermes #:-1073760704 (AppId:260001512) (line:111). 1095852103 5 1 Wed Sep 22 11:21:43 2004 260100138 tgp tgp_subsys scGantryInMail.c 725 Firmware SHUTDOWN occurred caused by this Ermes #:-1073760704 (AppId:260001512) (line:111).

Error: 260100137 Host caused the firmware subsystems to SHUTDOWN. See long description for full error Mon May 26 07:00:32 2008 Error: 260100137 Host: tgp Process: tgp_subsys File: scHostInMail.c Line: 438 Wed Dec 31 19:00:03 1969 Host : Table Gantry Processor Ermes # : 260100137 Exception Class : Secondary Severity : Sec/Soft File : scHostInMail.c Line# : 438 Function : No System Function Reported Scan Type : None/Unknown Scan: 0/0/0 Host caused the firmware subsystems to SHUTDOWN.

Applies to

Models
LightSpeed
Product
LightSpeed 5.X (80/100kW) LightSpeed 5.X (80/100kW) ~ Use with both MDAS and GDAS ~ GRE Console HP_H_G16 HP_H_M16 HP_O_M16 LightSpeed 5.X (100kW) GDAS16 with GRE Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
5370000 - TGPHD BOARD ASSY - TGPU For HD, HP and VCT (Replaces the 5271005-2 & 2349697-2, -3, -4 and -5). See Long Description about backward compatability. Part #: 5271005-2 Part Type: GP CONSUMABLES Part Weight: 0.00 LBR Repairable: No Returnable: No Hazardous Material: N Description: TGPHD BOARD ASSY Positioning VCTHD FRU: Yes Status: Orderable Comments: Per ECR 2098435, 212008898, 30 April 2010. Old FRU Compatible Yes Comments Per ECR 2098435, 212008898, 21 Jul 2010. Additional FRU Needed No Old FRU Scrapped No Old FRU Upgradable No Use on Hands Yes Use Demand History Yes Ultimate Prime 5271005-2 New FRU Backward Compatible Yes Old FRU Usable Yes Change Type Indirect ECO Number ECR2098435 2349697-5 is replacing the -4 version of this board. -3 versions are still in stock. If you are replacing parts in a VCT or HP gantry order the -3 until stock is depleted. ** FE ASSIST ** The 2349697-3 version is for VCT and all 5.x systems i.e. HP60 HP80- HP100's. The 2349697-4 version will works for CT750-HD and is backwards compatible but the -3 will not work on a CT750HD. (212001261 12/11/08) Part #: 2349697-5 Part Type: GP CONSUMABLES Part Weight: 0.00 LBR Repairable: No Returnable: No Hazardous Material: N Description: TGPU BD ASSY Positioning VCTHD FRU: Yes Status: Orderable Part #: 2349697-4 Part Type: GP CONSUMABLE - RETURNABLE Part Weight: 0.00 LBR Repairable: No Returnable: Yes Hazardous Material: N Description: TGPU BD ASSY Positioning HD FRU: Yes Status: Orderable Comments: ** FE ASSIST ** The 2349697-4 is for CT750-HD and is backwards compatible to The 2349697-3 used on VCT and all 5.x systems i.e. HP60 HP80- HP100's. Do not use the 2349697-3 on CT750HD (212001261 12/11/08) TGPU 2349697-3 Part #: 2349697-3 Part Type: GP CONSUMABLES Part Weight: 0.00 LBR Repairable: No Returnable: No Hazardous Material: N Description: TGPU BD ASSY Positioning HP60 FRU: Yes Status: Orderable Comments: Added per VCT RSPL, this is replacement for -2, and non dash versions. ms13452 050325
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.