Pular para o conteúdo principal
GE Healthcare · MRSigna20000238110628200002357

Cannot Boot Applications with Host Monitor Timeout Error Due to Corrupted config file(s).Cannot Boot Applications with H…

Cannot Boot Applications with Host Monitor Timeout Error Due to Corrupted config file(s).Cannot Boot Applications with Host Monitor Timeout Error.Do a restoreINFO and Then Reboot wqr.Signa HDxT 15.0

Symptom

Cannot Boot Applications with Host Monitor Timeout Error

Error message

200002381 Process scn, pid XXXX terminated by unexpected signal XX Tue Apr 22 12:15:11 2008 Error: 200002381 Host: HDXNMR Process: hostMonitor File: ssSigHandlers.c Line: 164 Exception Class: Unknown Function: Host : System Startup Process scn, pid 10628 terminated by unexpected signal 11

HostMonitor is not running after 400 seconds. System startup failed

200002357 A HOST_SHUTDOWN was initiated by the death of protocol since other processes depend on it to come up. 200002357 ANGEMRI hostMonitor ssProcStop.c 977 Exception Class: Unknown Function: Host : System Startup A HOST_SHUTDOWN was initiated by the death of protocol since other processes depend on it to come up.

Applies to

Models
Signa
Product
Signa HDxT This product is for 1.5T Signa HDxT systems with 15.0 software and HDi systems with 15.0 software. This product is for 1.5T Signa HDxT systems with 15.0 software and HDi systems with 15.0 software.
Software
15.0 For Product Signa HDxT and HDi only. -- NOT to be used with HDe. -- For Product Signa HDxT and HDi only. -- NOT to be used with HDe. --
Area
MR

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required