Symptom

One Gantry Control Panel says "OFF" and buttons will not respond. See long description for details Either the Left or Right Gantry Control Panel says "OFF" and the buttons will not respond. Panel appears to be bad.

Error message

Error: 182880 Node is Unreliable and has been SHUTDOWN. See long description for an example of error message. Wed Sep 8 08:15:54 2010 Error: 182880 Host: ETC Process: Gantry Display File: gd tnc erlog.c Line: 599 Function: No System Function Reported Scan: 6518/5/0 Type: None/Unknown Exception Level: Pri/Soft Time: 08:15:54:690 Log Series:47 Node is Unreliable and has been SHUTDOWN. Reason: GUIF CAN Node Has Reset Too Many Times in a Short Period of Time. Node Id :5 , Node Type: GANTRY CONTROL. Board Number: 2256729 Serial Number: 8639.

Error: 182869 GUIF CAN Gantry Control Timed Out on Updating its Button Status to the TNC. See long description for details Mon Feb 7 06:20:25 2005 Error: 182869 Host: ETC Process: Gantry Display File: gd_tnc_erlog.c Line: 471 Function: No System Function Reported Scan: 0/0/0 Type: None/Unknown Exception Level: Pri/Soft Time: 06:20:25:645 Log Series: 32 GUIF CAN Gantry Control Timed Out on Updating its Button Status to the TNC. Node Id: 5. Board Number: 0 Serial Number: 0.

Error: 182883 An Invalid RESET has been Detected. Node Type: TNC (ETC I/F). See Long Description for more details Mon May 9 13:45:23 2005 Error: 182883 ETC Gantry Display gd tnc 

Applies to

Models
LightSpeed
Product
LightSpeed 4.X - GRE See Long Description for detailed information - H2_P_M16 Product Name: LightSpeed 4.X with GRE Product Photographs GE HEALTHCARE Method of remote or software identification - See attachment below for a complete list of software commands and typical results 1. check_config 2 . swhwinfo “swhwinfo” will have the following switches: · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) -h or –help = Outputs complete details about all the available switches associated with swhwinfo swhwinfo –c –m –o -da List of Supported Console types: GOC3 (GRE) GOC4 (GRE) List of software the product can run: 07MW11.10 SP2.0, SP3.0, SP3.1 FAQ's Table weight limit - 180 kg (400 lbs) normal operation - 205 kg (450 lbs) maximum allowed with decreased accuracy Tube type - D3186T D3186T(green) Current production - M4 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
5310001-3 Left CT Control Panel - White N9 Bezel and Blue LEDs See Long Description Part #: 5310001-3Part Type: GP CONSUMABLES Part Weight: 0.00 LBRRepairable: NoReturnable: NoHazardous Material: NDescription: Left CT Control Panel for HDCT, with N9 Bezel and Blue LEDsFRU: YesStatus: Orderable
Area
CT

Full cause & step-by-step resolution