GE Healthcare · MRSigna2252133
Defective Patient Handling Power Supply Module (PHPS).
Symptom
Fails TPS Reset when booting up to Applications or when Resetting TPS See Long Description When the MR system is booting up to applications it will fail TPS reset but Applications will successfully boot. With Applications up and "Reset TPS" has been selected, TPS fails to download successfully.
Error message
2252133 SDO Communication Failure was detected on the CANopen Protocol Node Detecting Failure = SCP Wed Aug 25 08:04:47 2004 2252133 2 MR-OC0 NSP SCP:sdo_segment_upload 1408 SDO Communication Failure was detected on the CANopen Protocol. Node Detecting Failure = SCP. Node ID = 4. Index = 0x3003. SubIndex = 0x0. Error Code = 0x1002.->
Applies to
- Models
- Signa
- Product
- Signa HDxT This product is for 1.5T Signa HDxT systems with 15.0 software and HDi systems with 15.0 software. This product is for 1.5T Signa HDxT systems with 15.0 software and HDi systems with 15.0 software.
- Software
- 15.0 For Product Signa HDxT and HDi only. -- NOT to be used with HDe. -- For Product Signa HDxT and HDi only. -- NOT to be used with HDe. --
- Component
- 5215012 MR Patient Handling Power Supply see long description 5215012 MR Patient handling power supply for assembly for MR excite 3. This is identical to the 5109368 assembly but has power supplies with a modification to allow operation upto a 200 Gauss magnetic field.
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- Running grafidy3, and find there are spikes on the waveforms in the manual prescan display, both windows.
- Fails to Scan with TORSO Coil.
- The Carriage Fails to Advance to Latch to the Table and to Retract When it is Raised, Undocked or Lowered.
- AE title has changed on destination host.
- CWA ID100315. Auto and Manual Prescan Fails.
- What is the Acoustic Noise Level in the Ovation Magnet.