Pular para o conteúdo principal
GE Healthcare · MRSigna22522472252232

The CAN Link is Reporting Hard or Intermittent CAN Link Failures due to CAM Chassis.

Symptom

The CAN Link is Reporting Hard or Intermittent CAN Link Failures. Any device on the CAN Link may suddenly start reporting hard or intermittent CAN Link failures and CAN Link communications may be interrupted. Replacing the suspect part will not fix the problem.

Error message

2252247 CAN Network Down. Network Management Failures were detected on all CAN nodes. The whole network is down. see long description Thu May 20 15:24:59 2010 Error: 2252247 Host: MR1BOC Process: NSP File: SCP:CanLinkHWO Line: 1067 Exception Class: Unknown Severity: Unknown Function: Peripherals: Communications PSD: Unknown Coil: Unknown Scan: -1/-1 CAN Network Down. Network Management Failures were detected on all CAN nodes. The whole network is down.-> Extended Error: All of the CAN nodes stopped responding to watchdog messages from the SCP. Check that the Driver Module has power and that the CAN cables are connected to the Driver Module and SCP.This message is expected if the Driver Module is manually powered off. Possible cause(s): CAN Communication

2252232 SCP detected a Node Reset on the CAN network. Reset Node is %b.

Applies to

Models
Signa
Product
Signa HDxT This product is for 1.5T Signa HDxT systems with 15.0 software and HDi systems with 15.0 software. This product is for 1.5T Signa HDxT systems with 15.0 software and HDi systems with 15.0 software.
Software
15.0 For Product Signa HDxT and HDi only. -- NOT to be used with HDe. -- For Product Signa HDxT and HDi only. -- NOT to be used with HDe. --
Component
5181495-4 CAM Metal Cage with Mid plane 5181495-4 CAM Metal Cage with Mid plane
Area
MR