GE Healthcare · CTLightSpeed2601010232886126010110126010101250060260001514
LightSpeed RT16 & RT16 Xtra and Vision
Symptom
Attention: The Scanning Hardware is not available. Please see long description. Attention: The Scanning Hardware is not available, please wait for the hardware reset successfull status posted in the feature status area on the right monitor. If the hardware reset fails, then reset the scan hardware from the Service Desktop System Resets button.
Error message
Error: 260101023 The CCB has notified the system of an error condition which prevents scanning Thu Nov 20 10:38:27 2008 Error: 260101023 Host: orp Process: orp_subsys File: rcib_erlog.c Line: 463 Thu Nov 20 10:38:26 2008 Host : Collimator Ermes # : 260101023 Exception Class : Abort Severity : Primary File : rcib_erlog.c Line# : 463 Function : No System Function Reported Scan Type : Scout Scan: 28861/1/1 The CCB has notified the system of an error condition which prevents scanning. See log for further information Error: 260101101 The DCB has unexpectantly reset to the boot level. See long description for a full sample error. Tue Aug 31 08:50:57 2010 Error: 260101101 Host: orp Process: orp_subsys File: rcib_node_init.c Line: 392 Exception Class : Secondary Severity : Sec/Soft Function : No System Function Reported Scan Type : None/Unknown Scan: 0/0/0 The DCB has unexpectantly reset to the boot level. A Reset of the sub-system will be performed to attempt to recover from this error. Error: 260101012 The Rotating Controller Interface Bus (RCIB) fault line has been asserted. See LD for more details Mon Apr 10 07:20:41 2006 Error: 260101012 Host: orp Process: orp subsys Mon Apr 10 07:2
Applies to
- Models
- LightSpeed
- Product
- LightSpeed RT16 & RT16 Xtra and Vision WB_H_G16_G_GTL = RT16, Xtra WB_H_G16_G_GTL650 = Vision WB_H_G16_G_GTM = RT16, Xtra or Vision LightSpeed RT16 & RT16 Xtra with GRE and GDAS16 gantry Links to the Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Component
- 2361853-2 GDAS DCB 16-SLICE replaces 2361853 See LD for more Info 2361853-2 GDAS DCB 16-SLICE replaces 2361853 Part #: 2361853-2 Part Type: GP CONSUMABLES Part Weight: -59999.40 LBR Repairable: No Returnable: No Hazardous Material: N Description: GDAS DCB FRU: Yes Status: Orderable FLS - GDAS DCB (16-SLICE) USED ON: LIGHTSPEED PRO 16 GDAS ADDED PER GDAS RSPL MS13452 This board has been replaced by 2361853-2, FE NOTE: Site MUST have M4FMI completed before this board will function properly, ms13452
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required