table cannot move. c1621:reconnect the power cable on the module2 mother board the voltage rise to 5.15V. adjust the +5V…
table cannot move. c1621:reconnect the power cable on the module2 mother board the voltage rise to 5.15V. adjust the +5V power supply get 5.05V.
Symptom
prestige 2 table cannot move. c1621
Error message
error 1621, power supply under 5V
Applies to
- Product
- Prestige II
- Area
- X-ray R&F
Full cause & step-by-step resolution