Pular para o conteúdo principal
GE Healthcare · CTLightSpeedVCT244634260118327260118315

LightSpeed 7.x (VCT, VCT Select, Pro32)

Symptom

Scanner Hardware Stopped Scan. Scanner Hardware Stopped Scan

Error message

Error: 244634 Subsystem State; DCB:up, CCB:up, TCB:up, --- JEDI:down --- TGP/ORP:up Jedi Down See Long Description for details Wed Feb 17 23:30:55 2010 Error: 244634 Host: bay46 Process: fwmgr File: fwmgr_utils.c Line: 677 Function: No system function reported Scan Type: Not scanning Scan: 0/0/0 Scan Seq Id: 0 Subsystem State; DCB:up CCB:up TCB:up JEDI:down TGP/ORP:up

Error: 260118327 Can not initialize Jedi generator See Long Description for complete error code listing. Fri Oct 20 19:11:03 2006 Error: 260118327 Host: orp Process: orp_subsys File: gen_exec.c Line: 1122 Fri Oct 20 19:11:02 2006 Host : On-Board Rotating Processor Ermes # : 260118327 Exception Class : Abort Severity : Secondary File : gen_exec.c Line# : 1122 Function : No System Function Reported Scan Type : Helical Scan: 8810/2/2 Can not initialize Jedi generator If this continues to appear, generator configration is mismatched, or power and commnication are not interconnected.

Error: 260118315 JEDI ID Communication Timer Expired. Check power and commnication interconnects. See Long Description for complete error Mon Dec 6 12:04:49 2004 Error: 260118315 Host: orp Process: orp_subsys File: gen_exec.c Line: 1665 Mon 

Applies to

Models
LightSpeed, VCT
Product
LightSpeed 7.x (VCT, VCT Select, Pro32) V40_H_V64_G_GTM V40_H_V64_G_GTL V40_H_V32_G_GTM V40_H_V32_G_GTL V40_H_H64_G_GTM V40_H_H64_G_GTL V40_H_HS64_G_G Product Name: LightSpeed 7.x (VCT, VCT Select, Pro32) Product photographs Method of remote or software identification - See attachment for a complete list of software commands and typical results. check_config swhwinfo “swhwinfo” will have the following new switches available with the VCT M4 and PF 1.5 M3 software release (software release deploying this “swhwinfo” capability is “06MW03.?”): · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) · -h or –help = Outputs complete details about all the available switches associated with swhwinfo · swhwinfo –c –m –o -da List of Supported Console Types GOC5 Will have a VDARC and VIG GOC6 Will have a VDIG List of softwares the product can run 09MW08.10 09MW08.11 08MW33.2 07MW18.4 06MW03.4 05MW31.6 FAQ's Table weight limit - Up to 500 lb. (227 kg) with ± .25 mm positional accuracy Tube type - D3194T D3193T(green) Current production - M4 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system. Done acording to template
Component
5136161-4 PPC KV Control replaces (5136161, 5136161-2, 5117564-10) used by all 5.X, Brightspeed and VCT gantries See Long Description for Details PPC KV Control Bd. 5136161-4 replaces (5136161, 5136161-2, 5117564-10) this is the orderable Programmed KV Control Board for HP60, LSPro 80KW , 100KW and VCT 7.X, RT16, BrightSpeed Power and Select Systems (Velocity, Delight, Meri) Part #: 5136161-4 Part Type: GP CONSUMABLES Part Weight: 0.00 KGM Repairable: No Returnable: No Hazardous Material: N Description: PROGRAMMED kV CONTROL P6.33BF6 AS FRU FRU: Yes Status: Orderable Current: 5136161-4 Harvested: Not Available Exchange: Not Available Comments: Jedi KV Control Board used on Definium 8000, Revolution XR/d-2X (Lightning), Definium 6000, and Precision 500D. Also may be used on all Innova systems Bolus/Unity/SPAD/Harmony. P.Liebau 1/2010 Part #: 5136161-2 Part Type: GP CONSUMABLES Part Weight: 0.00 LBR Repairable: No Returnable: No Hazardous Material: N Description: PROGRAMMED kV CONTROL 6.30BFA AS FRU FRU: Yes Status: Orderable Comments: In order to avoid Jedi software download issues on Definium 8000, Revolution XR/d-2X (Lighting), Definium 6000, and Precision 500D you may order 5136161-4, KV Control Board developed for these systems. Part #: 5136161 Part Type: GP CONSUMABLES Part Weight: 2.43 LBR Repairable: No Returnable: No Hazardous Material: N Description: PROGRAMMED KV CTRL P6.30BFA as FRU FRU: Yes Status: Orderable Comments: (2/21/06 212001261) The 5136161 KV board replacement is the preferred part used on all HP60 hp 80 LS-RT and VCT products.
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.