Symptom
Reset Image Database did not run correctly and after console restart having image database errors.
Error message
Error: 9017 system error : No such file or directory See long description for a full sample error. Thu Sep 22 10:54:34 2011 Error: 9017 Host: ent_ct Process: imserver File: diskmisc.c Line: 230 system error : /usr/g/sdc_image_pool2/reserved/.IMSsetmwfCx No such file or directory Error: 1016 Unmatched object in database (system will correct): 0 - getobjectfilepath failed See long description for full error Fri Nov 21 06:51:37 2008 Error: 1016 Host: sdghct Process: dbwserver File: exec.ec Line: 488 Unmatched object in database (system will correct): 0 - getobjectfilepath failed
Applies to
- Models
- LightSpeed, VCT
- Product
- LightSpeed 3.X (Ultra, Plus, QX/i) - GOC2 This is HP CPU host running LINUX with Pegasus IG. 308L.8_H3.3M4 ~ 4 or 8 slice MDAS. ~ Could be an Ultra, Plus or QXi system. Must run swokinstall -p to verify. ~ See Long description for details . LightSpeed 3.X MDAS4 Gantry and GOC2 Gantry - 2281177-2, DAS - 2272768, Console – 2341104-2, Table - 2276776, PDU - 2269902 Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Component
- 2368697 Linux DDR Workstation - Used on GOC2 only See long description for details Part #: 2368697 Part Type: GP CONSUMABLE - RETURNABLE Part Weight: 0.00 LBR Repairable: No Returnable: Yes Hazardous Material: N Description: LINUX DDR WORKSTATION - GEMS CT USA FRU: Yes Status: Orderable Comments: FLS 3/5/05 (212001261 Shawn Fox) HOST COMPUTER (HP). THIS IS THE GOC2 HP 8000 COMPUTER . ***EXTREMELY IMPORTANT: NEW SOFTWARE, S/W LOAD PROCEDURE AND S/W OPTION KEYS ARE REQUIRED IF THE HP8000 COMPUTER IS REPLACED. Contact the OLC to assist with new Option Keys.
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.