Symptom

Cannot display several Images in Exam - document in RFS - critical issue Cannot display several Images in Exam

Error message

Error: 200002745 Stat failure after image closed See LD for details Tue Jun 7 12:27:18 2005 Error: 200002745 hostid: su101 process: imagecreate ic_imagecreate_req.c 536 Stat failure after image closed Image Path : /usr/g/sdc_image_pool2/reserved/IMSse802zHG/E302U0_S2U0_I7U0.11 18165232_474766 File stat status : -1 Errno : 5

Error: 2052 error during execution of the fonction (note: may or may not be misspelled) See LD for details Tue Jun 7 12:27:18 2005 Error: 2052 hostid: su101 process: imserver imsprim.c 595 error during execution of the fonction /usr/g/sdc_image_pool2/reserved Input/output error Note: fonction is not mis-spelled Wed Sep 12 10:26:07 2007 Error: 2052 Host: gect Process: imserver File: imsprim.c Line: 622 error during execution of the function /usr/g/sdc_image_pool/reserved Input/output error

Applies to

Models
LightSpeed, VCT
Product
LightSpeed 4.X - GRE See Long Description for detailed information - H2_P_M16 Product Name: LightSpeed 4.X with GRE Product Photographs GE HEALTHCARE Method of remote or software identification - See attachment below for a complete list of software commands and typical results 1. check_config 2 . swhwinfo “swhwinfo” will have the following switches: · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) -h or –help = Outputs complete details about all the available switches associated with swhwinfo swhwinfo –c –m –o -da List of Supported Console types: GOC3 (GRE) GOC4 (GRE) List of software the product can run: 07MW11.10 SP2.0, SP3.0, SP3.1 FAQ's Table weight limit - 180 kg (400 lbs) normal operation - 205 kg (450 lbs) maximum allowed with decreased accuracy Tube type - D3186T D3186T(green) Current production - M4 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
2368697 Linux DDR Workstation - Used on GOC2 only See long description for details Part #: 2368697 Part Type: GP CONSUMABLE - RETURNABLE Part Weight: 0.00 LBR Repairable: No Returnable: Yes Hazardous Material: N Description: LINUX DDR WORKSTATION - GEMS CT USA FRU: Yes Status: Orderable Comments: FLS 3/5/05 (212001261 Shawn Fox) HOST COMPUTER (HP). THIS IS THE GOC2 HP 8000 COMPUTER . ***EXTREMELY IMPORTANT: NEW SOFTWARE, S/W LOAD PROCEDURE AND S/W OPTION KEYS ARE REQUIRED IF THE HP8000 COMPUTER IS REPLACED. Contact the OLC to assist with new Option Keys.
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.