GE Healthcare · CTLightSpeed230017812000500230017801
LightSpeed 5.X (60kW)
Symptom
Single channel ring artifact on Axial images (partial ring on Helical) with Tangential streaks on outside of ring. See LD for artifact pictures Single channel ring artifact on Axial images (partial ring on Helical) with Tangential streaks on outside of ring.
Error message
Error: 230017812 ACCS CCal: Error: too much variation in ratio data. Limit 0.000500 See Long Desciption for complete error Mon May 30 05:41:49 2005 Error: 230017812 Host: cxi1 Process: accs File: collSweepProcessing.c Line: 491 ACCS CCal: Error: too much variation in ratio data. Limit 0.000500 Error: 230017801 ACCS CCal: Sweep Processing failed. Status = -1 See the Long Description for the full error code Tue Nov 30 02:33:04 2004 Error: 230017801 Host: k2_ct2_ls Process: accs File: ccal_prep.c Line: 661 ACCS CCal: Sweep Processing failed. Status = -1
Applies to
- Models
- LightSpeed
- Product
- LightSpeed 5.X (60kW) LightSpeed 5.X (60kW) ~ Use with MDAS or GDAS HP_P_G4 HP_P_G8 HP_P_G16 HP_P_M16 LightSpeed 5.X 60KW G-DAS16 with GRE Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Component
- 2404393 GDAS-16 CONVERTOR CARD, replaces 2364354. See Long Description for details. 2404393 GDAS-16 CONVERTER CARD, replaces 2364354 Part #: 2404393 Part Type: GP CONSUMABLES Part Weight: 2.20 LBR Repairable: No Returnable: No Hazardous Material: N Description: 16 SLICE GDAS CONVERTER CARD FRU, A CONVERTER CARD IN AN ESD BAG IN A BOX WITH A PART NUMBER LABEL FRU: Yes Status: Orderable Comments: FLS - 16 SLICE GDAS CONVERTER CARD USED ON: LIGHTSPEED PRO 16 FRU GDAS ADDED PER GDAS RSPL MS13452 040315 RISK: SITE DOWN FREQUENCY: HIGH NOTE: THERE ARE 96 OF THESE CARDS PER SYSTEM, BUT SHOULD NOT NEED TO REPLACE ALL AT ONETIME, MS13452 THIS PART CONTAINS ONE GDAS CONVERTER CARD 2364354 WITH FRU PACKAGING, SUPPLIED BY GEMS IT. MS13452 PDR 8471.80.9000 IM05 UPDATE AWAITING MM (212001261 4/1/06) For 4/8 slice G Das which uses a thick connector must order 2364354-2
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required