Pular para o conteúdo principal
GE Healthcare · CTLightSpeed178573

LightSpeed 3.X (Ultra, Plus, QX/i) - Octane

Symptom

Scanner Hardware Reset Failed. See LD Scanner Hardware Reset Failed Hardware Failed to reset message appeared on monitor AFTER boot-up WITH Error:178568 Failed to establish socket connection to STC in 10 seconds and errors 244553 and 244554 and power up self test fails A, B, and 9 on STC CPU board The following errors were in the gesyslog Thu Jan 13 04:18:32 2005 Error: 178568 Suite: RP25 Host: SBC0 Process: fwmgr File: NBSSocket.m 1.9 Method: connectToTarget Line: 315 Failed to establish socket connection to STC in 10 seconds Thu Jan 13 04:18:32 2005 Error: 244553 Suite: RP25 Host: SBC0 Process: fwmgr File: HwSetUp.m 1.18 Method: hardReset Line: 429 The STC controller has failed to reset. Thu Jan 13 04:18:32 2005 Error: 244554 Suite: RP25 Host: SBC0 Process: fwmgr File: HwSetUp.m 1.18 Method: hardReset Line: 442 The reset failure of the OBC may be due to the reset failure of the STC. Also saw failures A, B, and 9 on the power-up test on the STC CPU .

Error message

Error: 178573 The following controllers reported a failure when initializing: OBC/ETC/STC See Long Description for complete error Sun Aug 6 08:41:24 2006 Error: 178573 Host: coxsct2 Process: fwmgr File: HwSetUp.c Line: 2005 Function: No system function reported Scan Type: Not scanning Scan: 0/0/0 Scan Seq Id: 0 The following controllers reported a failure when initializing: OBC/ETC/STC

Applies to

Models
LightSpeed
Product
LightSpeed 3.X (Ultra, Plus, QX/i) - Octane Always 4 or 8 slice MDAS 308i.2_H3.1M5 LightSpeed 3.X - Octane Method of visual identification Method of remote or software identification - See attachment below for a complete list of software commands and typical results. check_config List of Supported Console Types Octane with Pegasus Recon hardware. List of softwares the product can run 308I.2_H3.1M5 305.4_H3.1M4 - obsolete 303.1_H3M4 - obsolete FAQ's Table weight limit - Up to 400 lb. (180 kg) with ± .25 mm positional accuracy. 450Ib (205 kg) max weight Tube type - D3186T Current production details - See Technical Reference Manual found on the Common Document Library Tips and workarounds, where to find it - Common Document Library(CDL) . Search for the latest FMI software release. In the furnished materials will be the list of various manuals with part numbers for the various languages. Example: English versions 2211223-100 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
2224296-3 Artysen board for 1X/2X/3X/4X used in the ETC/STC/OBC See Long Description for FEMC information and board locations Part #: 2224296-3 Part Type: GP REPAIRABLES Part Weight: 1.09 LBR Repairable: Yes Returnable: Yes Hazardous Material: N Description: VE-GE, 8MB, Artesyn CPU Bd FRU: Yes Status: Orderable Parts Offered as: Current : 2224296-3 Harvested : Not Available Exchange : Not Available Comments: ** FE ASSIST ** Repair operations has found that 68% of the time the Artesyn board (2224296-3) is returned for repair, it has No Defect Found. Check switch settings (212001261 2/12/09)
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required

Related GE Healthcare CT issues