GE Healthcare · CTLightSpeedVCT200352906290009008290009030
LightSpeed 1.X - GRE
Symptom
Recon self test failure on boot up. Recon self test failure on boot up
Error message
DARC /var/log/messages : timeout attempting to rsh to ig2 ( can be ig1 or ig3 ) See long description DARC /var/log/messages Aug 18 18:43:09 darc logger: is_IG_up 1 ig2 called Aug 18 18:43:09 darc is_IG_up:[recon_control:7973]: ig3 is up ( this will be typically 5 mins after start of DARC boot ) Aug 18 18:43:09 darc is_IG_up:[recon_control:7967]: ig1 is up Aug 18 18:43:10 darc is_IG_up:[recon_control:7970]: timeout attempting to rsh to ig2 ( this is ig2 failing to respond to DARC ) With no intervention DARC will retry to contact ig that failed to respond on fisrt attempt ( notice timestamps ) Aug 18 18:49:58 darc logger: is_IG_up 3 ig2 called Aug 18 18:49:58 darc mountd[3273]: authenticated mount request from ig2:855 for /usr/g (/usr/g) Aug 18 18:49:58 darc is_IG_up:[recon_control:7970]: ig2 is up ( ig2 up nearly 7 minutes after second start ) Error: 200352906 Errors starting up the recon sub system. See Long Description for complete error Mon Oct 10 11:36:18 2005 Error: 200352906 Host: wbvct1 Process: reconSelftest File reconSelftest.c Line: 314 Errors starting up the recon sub system. Status = 1237 Error: 290009008 Timeout waiting for Image Set Complete from IG3. (could be IG1 o
Applies to
- Models
- LightSpeed, VCT
- Product
- LightSpeed 1.X - GRE H1_P_S4 Lightspeed 1.X - GRE Method of visual identification Large square gantry similar in appearance to a CTi gantry. Method of remote or software identification - See attachment below for a complete list of software commands and typical results. check_config “swhwinfo” will have the following new switches available with the VCT M4 and PF 1.5 M3 software release (software release deploying this “swhwinfo” capability is “06MW03.?”): · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) -h or –help = Outputs complete details about all the available switches associated with swhwinfo swhwinfo –c –m –o -da List of Supported Console Types GOC3 Will have a DARC, SDDA and IG GOC4 Will have a DARC 2 and IG List of softwares the product can run 07MW11.10 SP3 06MW29.7 - obsolete 06MW03.5 - obsolete 05MW14.5 - obsolete 04MW44.9 - obsolete 04MW13.7 - obsolete FAQ's Table weight limit - Up to 400 lb. (180 kg) with ± .25 mm positional accuracy. 450Ib (205 kg) max weight Tube type - D3176T Current production details - See Technical Reference Manual found on the Common Document Library Tips and workarounds, where to find it - Common Document Library(CDL) . Search for the latest FMI software release. In the furnished materials will be the list of various manuals with part numbers for the various languages. Example: English versions 2211223-100 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Component
- 5147443-2 Jarrell IG for NON VCT systems (replaces 5147443 ). See long description for details Part #: 5147443-2 Part Type: GP CONSUMABLE - RETURNABLE Part Weight: 0.00 LBR Repairable: No Returnable: Yes Hazardous Material: N Description: Jarrell GRE IG with RAC rev.D or Higher FRU: Yes Status: Orderable Comments: Per ECR 2040301 (Part of Jarrell NPI), 212008898, 3/8/07. **FE Assist** applies to CT but not PET systems. Must order software release 07MW11.10, p/n 5220443 (deployed by FMI 25389), and perform a LFC to support this FRU if system is not currently running sw rls 06MW29.7 or 07MW11.10 (8/16/07 212001261) 212008041 12/03/2007 15:23 GMT Part #: 5147443 Part Type: GP CONSUMABLE - RETURNABLE Part Weight: 0.00 LBR Repairable: No Returnable: Yes Hazardous Material: N Description: Computer - Image Generation, 2nd Generation FRU: Yes Status: Orderable Comments: Per Jarrell RSPL, 7/19/06, ms11892. ** FE Assist ** If ordering this Jarrell IG to replace a Westville IG 2362872 must also order a 5127222-2 Shielded Network cable and replace the existing one on your system Plus, must order software release 06MW29.7, p/n Part #: 5193989 and perform a LFC. (12/8/05 212001261
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.