A noise is heard only during gantry rotation - document in RFS - critical issue
Symptom
A grinding, rumbling, squealing, clicking, or other noise is heard only during gantry rotation - document in RFS - critical issue see long description for more detail. A grinding,rumbling, squealing, clicking, or other noise is heard only during gantry rotation. Gantry speed or velocity errors are NOT reported to the customer or the gesyslog. System will still scan ok but customer just complains of excessive noise from gantry. Sound volume or quality can change with rotation speed.
Error message
Error: 260118251 (80-1403)Connectic Fault. One of the flat cable connector is not connected in Generator See the Long Description for the complete error Thu Dec 14 11:23:13 2006 Error: 260118251 Host: orp Process: orp subsys File: gen exec.c Line: 2376 Thu Dec 14 11:23:12 2006 Host : Jedi Ermes # : 260118251 Exception Class : Abort Severity : Primary File : gen exec.c Line# : 2376 Function : No System Function Reported Scan Type : None/Unknown Scan: 229/2/0 80-1403)Connectic Fault. One of the flat cable connector is not connected in Generator.^C Z6,Z7= 0h,0h
Applies to
- Models
- Discovery, LightSpeed
- Product
- Discovery CT750 HD V40_PS_HD64_G_GTL V40_PS_HD64_G_GTM Discovery HD CT750 Method of visual identification HD gantry: Gray top and Gray Side covers with HD Markings Console Type: GOC6 has black SCIM and tower face Method of remote or software identification - See attachment below for a complete list of software commands and typical results. check_config swhwinfo “swhwinfo” will have the following new switches available · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) -h or –help = Outputs complete details about all the available switches associated with swhwinfo swhwinfo –c –m –o -da List of softwares the product can run (Consult attached FCT Software Cross Reference): 08MW39.4 - M3 (see attachement for M3 available options) 09MW29.14 - M4 (HD Options = M3 plus Gemstone Spectral Imaging, Gemstone Spectral Review and volume Helical Shuttle ) Example: Of difference with options software [insite@ct01 ~]$ swokinstall -p Product Software Option Keys: Gemstone Spectral Imaging Permanent HelicalShuttle Permanent Xtream Recon Permanent FAQ's Table weight limit: 500 lbs Tube Type - D3702T Green D3701T Current production - M4 Refer to product data sheets or Technical reference Manual, English 5317222-1EN Revision : 2 or later Refer to Application Tips and Work-Arounds Direction 5311815-1EN Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Component
- 5316402 jedi SC V3 cable set. used on HD750 see long description Part #: 5316402 Part Type: GP CONSUMABLES Part Weight: 0.00 KGM Repairable: No Returnable: No Hazardous Material: N Description: Jedi SC V3 cables set - FRU FRU: Yes Status: OrderableParts Offered as: Current: 5316402 Harvested: Not Available Exchange: Not Available Comments: Per HDVCT RSPL, 212008898, 9/10/08.
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.