Defective Master Exciter2 board.
Symptom
Scan pauses during scan with Error 2254012 RRF-DIF Fiber Optic Link Error on the Master RRF-DIF board. 20 MHz Clock Present signal is in an Invalid State. See Long Description Scan pauses during scan. Customer can scan again after TPS Reset. Error shown below is logged. 2254012 RRF-DIF Fiber Optic Link Error on the Master RRF-DIF board. 20 MHz Clock Present signal is in an Invalid State.-> Extended Error : Check the Gigalink fiber optic connection between the IRF I/O and RRF-DIF Boards. If IT cables are connected and the system indicates that the 20MHz clock is missing then check the state of the voltage status LEDS on the rear of the Exciter Board. If any are not illuminated then check for missing voltages from the test points on the front of the RRF power supply modules.
Error message
2254012 RRF-DIF Fiber Optic Link Error on the Master RRF-DIF board. 20 MHz Clock Present signal is in an Invalid State.-> See Long Description Extended Error: Check the Gigalink fiber optic connection between the IRF I/O and RRF-DIF Boards. If IT cables are connected and the system indicates that the 20MHz clock is missing then check the state of the voltage status LEDS on the rear of the Exciter Board. If any are not illuminated then check for missing voltages from the test points on the front of the RRF power supply modules
Applies to
- Models
- Signa
- Product
- Signa HDxT This product is for 1.5T Signa HDxT systems with 15.0 software and HDi systems with 15.0 software. This product is for 1.5T Signa HDxT systems with 15.0 software and HDi systems with 15.0 software.
- Software
- 15.0 For Product Signa HDxT and HDi only. -- NOT to be used with HDe. -- For Product Signa HDxT and HDi only. -- NOT to be used with HDe. --
- Component
- 2361470 or 2361470-2 REMOTE RF MASTER EXCITER BOARD or RRF Master Exciter 2 Board with Temperature Compensation 2361470-3 2361470 or 2361470-2 REMOTE RF MASTER EXCITER BOARD or RRF Master Exciter 2 Board with Temperature Compensation 2361470-3
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- Running grafidy3, and find there are spikes on the waveforms in the manual prescan display, both windows.
- Fails to Scan with TORSO Coil.
- The Carriage Fails to Advance to Latch to the Table and to Retract When it is Raised, Undocked or Lowered.
- AE title has changed on destination host.
- CWA ID100315. Auto and Manual Prescan Fails.
- What is the Acoustic Noise Level in the Ovation Magnet.