Symptom

The positioning lights will not come on when the Patient was being positioned for a scan The laser lights will not come on when the Patient was being positioned for a scan

Error message

Error: 260006813 Gantry Failed to change to Commanded Position Velocity of 4.0 secs/rev within 15.000 seconds See LD for complete error Host: STC Proc: Axial Error: 260006813 File: amc_exec.c Line: 3326 Function: AXIAL CONTROL : axial control Scan 20/2/0 Type: None/unknown Exception Level: Pri/Most Severe Time: 17:59:36:615 Log Series: 405 Gantry Failed to change to Commanded Position Velocity of 4.0 secs/rev within 15.000 seconds. (AMC state:6 Cmd: 0x17d400b4 CurVel: 4179 (.001 sec/rev)) Retry Rx to verify probable axial drive or Gantry Balance or Encoder Hardware

Error: 260006815 Axial Motor Control (AMC) Logic could did NOT position the Gantry within tolerance. See long description for full error Mon Dec 6 09:27:53 2004 Error: 260006815 Host: STC Process: Axial File: amc_exec.c Line: 1494 Function: AXIAL CONTROL : axial control Scan: 23391/1/0 Type: None/Unknown Exception Level: Pri/Most Severe Time: 09:27:54:170 Log Series: 830 Axial Motor Control (AMC) Logic could did NOT position the Gantry within tolerance. Current Position: 108.7 degs Commanded Position: 90.0 degs Degrees Tol: +/- 1. Retry Rx to verify probable Gantry Balance or Brake or Encoder Hardware.

Applies to

Models
LightSpeed, VCT
Product
LightSpeed 4.X - Octane sw_version: 405I.2_H4.0M5 LightSpeed 4.X - Octane Gantry Model # 2339985 Console Model #'s 2341923 and 2341923-2(D3D) check_config shows LightSpeed 16 / LightSpeed 16 Product: LightSpeed16 sw_version: 405I.2_H4.0M5 or newer machine_type: Octane host_id: 3232297858 tube_type: 15 Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
5311936 Axial Encoder Assembly, Hub-Mounted Wheel. Used on H2, HP, VCT, and HD750. (Replaces 5182284-2 /5182284 / 2369747-2 / 2369747 / 2328307) See LD for details 5311936 Axial Encoder Assembly, Hub-Mounted Wheel. Used on H2, HP and VCT. (Replaces 5182284-2 /5182284 / 2369747-2 / 2369747 / 2328307) Part #: 5311936 Part Type: GP CONSUMABLES Part Weight: 0.00 LBR Repairable: No Returnable: No Hazardous Material: N Description: Shipping collector for Axial Encoder Assembly. FRU: Yes Status: Orderable Comments: Per ECR 2058601, 212008898, 10/8/08
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.