GE Healthcare · CTLightSpeed26011821310153
LightSpeed 5.X (60kW)
Symptom
Scanner Hardware Fault with high voltage spits. Scanner Hardware Fault with high voltage spits
Error message
Error: 260118213 (60-0316) Spit Max error. PPC board has detected the number of tube spits during exposure has reached the limit See Long Description for two error examples Sat Dec 17 18:32:18 2005 Error: 260118213 Host: orp Process: orp_subsys File: gen_exec.c Line: 2150 Sat Dec 17 18:32:17 2005 Host : Jedi Ermes # : 260118213 Exception Class : Abort Severity : Primary File : gen_exec.c Line# : 2150 Function : No System Function Reported Scan Type : Axial Scan: 10153/2/2 (60-0316)Spit Max error. PPC board has detected the number of tube spits during exposure has reached the limit (see theory of operation, software section).^C Z6,Z7= ch,0h
Applies to
- Models
- LightSpeed
- Product
- LightSpeed 5.X (60kW) LightSpeed 5.X (60kW) ~ Use with MDAS or GDAS HP_P_G4 HP_P_G8 HP_P_G16 HP_P_M16 LightSpeed 5.X 60KW G-DAS16 with GRE Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Component
- 5128609-7 (replaces 5128609) Anode HV CABLE Used on HP60 and BrightSpeed Systems. Introduced April 2006 to correct thermal failure of 2241313-2 P9230TG High Voltage Anode cable See Long Description for details 5128609-7 HV cable - Anode Part #: 5128609-7 Part Type: GP CONSUMABLES Part Weight: 0.00 LBR Repairable: No Returnable: No Hazardous Material: N Description: HV CABLE FRU: Yes Status: Orderable Old FRU Compatible No Comments Added per to ECR2115693 by 305024364 in April 7, 2011 Additional FRU Needed No Old FRU Scrapped No Old FRU Upgradable No Use on Hands Yes Use Demand History Yes Ultimate Prime 5128609-7 New FRU Backward Compatible Yes Old FRU Usable Yes Change Type Direct ECO Number ECR2115693 Part #: 5128609 Part Type: GP CONSUMABLES Part Weight: 0.00 LBR Repairable: No Returnable: No Hazardous Material: N Description: HV CABLE Positioning HP60 FRU: Yes Status: Orderable Comments: do not place new order for old p/n(2241313-2), existing old part should be used on HiSpeed Zx/i series only. D3186T PERFORMIX ULTRA TUBE UNIT Used on LS2.X, 3.X, 4.X, 5.X HP60, Hispeed QXi, DLS with H2 Gantry and DST with H2 Gantry. See Long Description for more info and Picture D3186T PERFORMIX ULTRA TUBE UNIT Used on LS2.X, 3.X, 4.X, 5.X HP60, Hispeed QXi, DLS with H2 Gantry and DST with H2 Gantry. Part #: D3186T Part Type: GP TUBES Part Weight: 184.31 LBR Repairable: No Returnable: Yes Hazardous Material: N Description: PERFORMIX ULTRA TUBE UNIT FRU: Yes Status: Orderable Comments: ** FE ASSIST ** The Shipping weight of this Tube is 336 lbs = 152 kilogram. For LightSpeed 2X/3X/4X and HP60. For CTi order D3132T and Lightspeed 1X order D3176T (212001261 5/3/07) *** SECOND OPINION NOTES *** Perform Rotor Control Diagnostics--If tube rotor fails to reach RUN mode in three ACCEL steps look for: High Drive output currents Low Drive output voltages and Low Drive output frequencies. This indicates a load problem with tube. With rotor running turn off Gantry 120V power at PDU. Listen for excessive noise and short deceleration time. A good rotor will coast for several minutes without power. Rotor issues (errors 183316, 183318, 183319, 260118111, 260118177) caused by tubes tend to be more prevalent on .5 second scans. If tube is installed on Jedi system, the Anode cable tends to break down (hard errors). Intermittent HV Spits -- Run KV Loop, 140KV, 30 sec, Anode, Cathode and both. Listen to tanks with stethoscope or screw driver, any noise indicates bad tank. Any spits? Remove HV cables from tanks fill with 2inch of oil. Spits = No, could be tube or cables. Spits = Yes, bad tank. D3186T PERFORMIX ULTRA TUBE UNIT Used on LS2.X, 3.X, 4.X, 5.X HP60, Hispeed QXi, DLS with H2 Gantry and DST with H2 Gantry. 2214527 Jedi standard HV tank for HiSpeed Xi, HP 60 and BrightSpeed See Long Description for details Part #: 2214527 Part Type: GP CONSUMABLES Part Weight: -3059969.40 LBR Repairable: No Returnable: No Hazardous Material: N Description: JEDI STD HV TANK (SPARE) FRU: Yes Status: Orderable Comments: FLS - JEDI STD HV TANK U/O YMS PROSPEED LXI,FXI,DXI REPLACES 2186730 DIFFERENCE IS PACKAGING FOR SERVICE. HQT: QUALITY ISSUE 10/98 ALL TANKS WITH SERIAL NUMBER PRIOR TO 22124BU4 ARE BAD. FE SHOULD CHECK HIS SERIAL # EUROPE PURGING GLOBE WILL SEND NEW MATERIAL AS QUICKLY AS POSSIBLE. AMERICA HAS NO STOCK ONHAND AS OF 10/22/98MS08992 RETURN DEFECTIVES TO: SWAP (MICHELL CHRISTOFFERSON), 2200 E. COLLEGE, BUILDING 11, CUDAHY, WI 53110. ADDED BY MS20544 ON 3/25/02. RETURN DEFECTIVES TO RECYCLING7/23/01 MS5277 TAKEN OFF RECYCLING PROGRAM 8/30/01 MS5277 TAKEN OFF RECYCLING PROGRAM PER L.BENIWAL (MSA8429 09/05/02) GELS - 2214527 = 2186730 + WOODEN BOX 10DISPUTES DEFECTIVE PARTS TO BE PUT ON SHELVES FOR ANALYSIS PDB00/07/27 KANBAN CSS F00,JEDI STD HV TANK (SPARE) 2214527 Jedi standard HV tank for HiSpeed and HP 60 Systems HV Tank for HP60 systems - must be disassembled from the Inverter Power assembly, and the new tank installed. Make certain you follow the disassembly/reassembly instructions in the Service Manuals exactly, as the process is quite difficult, and you are prone to scraped knuckles and possible broken connectors or scraped cables if you are not very careful. Picture of removed HV tank below:
- Area
- CT
Full cause & step-by-step resolution