GE Healthcare · CTLightSpeed2601210212050020316
LightSpeed 5.X (80/100kW)
Symptom
Scanner Hardware Stopped Scan. Scanner Hardware Stopped Scan
Error message
Error: 260121021 Tilt Stopping Error. Tilt did not reach target angle See LD for complete error Fri Dec 21 16:47:02 2007 Error: 260121021 Host: tgp Process: tgp_subsys File: tilt_monitor.c Line: 299 Fri Dec 21 16:47:01 2007 Host : Table Gantry Processor Ermes # : 260121021 Exception Class : Secondary Severity : Pri/Soft File : tilt_monitor.c Line# : 299 Function : No System Function Reported Scan Type : Helical Scan: 7649/2/1 Tilt Stopping Error. Tilt did not reach target angle -20500 mdeg. Tilt current angle: -20316 mdeg. Tilt angle tolerence: 250 mdeg. Timeout: 1000 msec.
Applies to
- Models
- LightSpeed
- Product
- LightSpeed 5.X (80/100kW) LightSpeed 5.X (80/100kW) ~ Use with both MDAS and GDAS ~ GRE Console HP_H_G16 HP_H_M16 HP_O_M16 LightSpeed 5.X (100kW) GDAS16 with GRE Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Component
- 46-136334P57 SPDT,MINIATURE BASIC SWITCH See long description. 46-136334P57 SPDT,MINIATURE BASIC SWITCH Part #: 46-136334P57 Part Type: GP CONSUMABLES Part Weight: 0.20 KGM Repairable: No Returnable: No Hazardous Material: N Description: SPDT, MINIATURE BASIC SWITCH.(Honeywell, V7 serie) FRU: Yes Status: Orderable IParts Offered as: Current: 46-136334P57 Harvested: Not Available Exchange: Not Available Comments: FLS - GRS PART **SLOW MOVING INVENTORY REPRICED PER SHAKUN 04/03**
- Area
- CT