Symptom

"Scan hardware Stopped Scan" with Axial Errors. "Scan hardware Stopped Scan" with Axial Errors

Error message

Error: 230023053 Failed on Gantry Firmware Mon Feb 4 07:19:52 2008 Error: 230023053 Host: aafgect1 Process: dailyPrep Rx File: DailyPrepUtils.c Line: 5433 Failed on Gantry Firmware. Cannot continue.

Error: 260107405 The Axial AC Drive reported fault code: F:5 Overvoltage Fault Mon Aug 14 07:14:49 2006 Error: 260107405 Host: tgp Process: tgp_subsys File: AX2fault.c Line: 191 Mon Aug 14 07:14:45 2006 Host : Table Gantry Processor Ermes # : 260107405 Exception Class : Abort Severity : Primary File : AX2fault.c Line# : 191 Function : AXIAL CONTROL : Axial Communications Scan Type : None/Unknown Scan: 0/0/0 The Axial AC Drive reported fault code: F:5 Overvoltage Fault The Drive internal DC bus voltage has exceeded 810V. Possible fail ure of Axial I/F Board Chopper Control or excessive motor regeneration from AXMotor during braking.

Applies to

Models
LightSpeed
Product
LightSpeed RT16 & RT16 Xtra and Vision WB_H_G16_G_GTL = RT16, Xtra WB_H_G16_G_GTL650 = Vision WB_H_G16_G_GTM = RT16, Xtra or Vision LightSpeed RT16 & RT16 Xtra with GRE and GDAS16 gantry Links to the Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
2233402 Axial Drive Motor Controller See Long Description for details Part #: 2233402 Part Type: GP CONSUMABLE - RETURNABLE Part Weight: 8.64 LBR Repairable: No Returnable: Yes Hazardous Material: N Description: AX DRIVE MTR CONTRLLER W/CORERS ASSY FRU: Yes Status: Orderable Comments: FLS - H2 GANTRY: MUST ORDER FROM HOLT ELECTRONICS. MS26561 DSM 1/7/01 USED ON HPOWER PROGRAM, PER RSPL. MS20544 ON 2/12/03. DO NOT DISCOUNT - PER SHAKUN*** 2ND OPINION NOTES *** Check fuses/connections associated with axial drive circuit; run axial functional diagnostic test and check for failures. If used on HP60
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.