Definium 6000 no X-ray exposure to wallstand detector, Error 200013 & 200012 The X-ray tube has reached prep s…
Definium 6000 no X-ray exposure to wallstand detector, Error 200013 & 200012
Symptom
Definium 6000 no X-ray exposure to wallstand detector, Error 200013 & 200012
The X-ray tube has reached prep state but fails to make any exposure to detector. A message “Exposure stopped. Release exposure switch and try again. Call service if problem persists.” is displayed on the monitor.
Error message
Error 200013: Time out while waiting for an event or an actor to respond. Error 200012: Time out while in WAIT_FOR_STOP_SEQ waiting for STOP_EXPOSURE_TRIGGER_DONE.
Applies to
- Models
- Definium
- Product
- Definium 6000
- Component
- 2375101 NEXT GENERATION APOLLO POWER SUPPLY
- Area
- X-ray R&F
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare X-ray R&F issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare X-ray R&F issues
- AEC Backup Time message
- Definium 8000 Auto positioning does not work after a reset/reboot due to wall stand receptor at a hard limit during rebooting or repowering.
- Definium 8000 One or both monitors display the following message on every boot
- Definium 8000 Wallstand or Table Only Configurations running VolumeRAD M3 Software.
- Definium 8000 OTS Collimator has no power due to a tripped breaker (CB8) ont he Tube Suspension Logic (TSL) Board.
- Definium 8000 The Siemens Collimator has not power due to a tripped breaker (CB9) on the Tube Suspension Logic (TSL) Board.