GE Healthcare · MRSigna222362122776622277672153735722776761525468

ICN disk error 2277662 in log Due To ICN Disk.

Symptom

ICN disk error 2277662 in log No issue seen by customer. 2277662 Disk error reported on ICN%s.

Error message

2223621 Internal Software Error. Can't callback on undefined NMRID:CM_ECHO_TPS. 2223621 Exception Class: Unknown Function: Software/Firmware: Host Communications Internal Software Error. Can't callback on undefined NMRID:CM_ECHO_TPS.

2277662 Disk error reported on ICN%s 2277662 ERROR: EM_DIAG_ICN_SMART_SUMMARY Disk error reported on ICN:%s Reported summary is as follows. Disk Device: %s Device String: %s Device Type: %s First Reported Date: %s Message: %s Extended Error: Run ICN Hard Disk Bandwidth test. If Bandwidth test fails: Stop scanning. Replace the ICN hard disk. Reconfigure ICNs. If Bandwidth test passes: Run ICN Hard Disk Integrity test If Integrity test passes: Continue scanning. Place order for new hard disk. Replace and reconfigure once new disk is obtained If Integrity test passes: Stop scanning. Replace hard disk immediately and reconfigure ICN. Possible cause(s): Img Compute Node, ICN Disk

2277672 TPS Reset ICN Boot Log Analysis detected an improper shutdown Mon Feb 23 06:44:04 2009 Error: 2277672 Host: EPMR02 Process: diagCoord File: icnBootLogCheck.pl Line: 358 Exception Class:Notification Function: Software/Firmware/Config:N/A TPS Reset (ICN Boot Log Analysis) d

Applies to

Models
Signa
Product
Signa HDxT This product is for 1.5T Signa HDxT systems with 15.0 software and HDi systems with 15.0 software. This product is for 1.5T Signa HDxT systems with 15.0 software and HDi systems with 15.0 software.
Software
15.0 For Product Signa HDxT and HDi only. -- NOT to be used with HDe. -- For Product Signa HDxT and HDi only. -- NOT to be used with HDe. --
Component
SecondOpinion Image Compute Node 5127452 See comments in Resolution Concept for second opinion requirements before ordering this part.
Area
MR