Pular para o conteúdo principal
GE Healthcare · CTLightSpeedVCT19435735258800501328656411849261671197311093

LightSpeed 7.x (VCT, VCT Select, Pro32)

Symptom

Slow Network of Images to a remote host. Slow Network of Images to PACS. It might also be not able to send images at all.

Error message

Network errors in netstat & low speed reported in ethtool (see LD) See long description. VCT Titan - Slow network eth2. {ctuser@ct01}[1] check_config Product: LightSpeed VCT sw_version: 10MW06.5 sw_hw_string: 10MW06.5_SP1-0-1.V40_H_V64_G_GTL host_id: 1943573525 tube_type: 16 gantry_type: GANTRY_VCT_40 No. of processors: 4 Processor speed: 2660 memory: 4096 {ctuser@ct01}[2] sudo dmidecode|grep -i hp Product Name: HP Z400 Workstation {ctuser@ct01}[3] {ctuser@ct01}[3] {ctuser@ct01}[3] {ctuser@ct01}[3] ifconfig eth0 Link encap:Ethernet HWaddr 00:26:55:D8:94:15 inet addr:172.16.0.1 Bcast:172.16.0.255 Mask:255.255.255.0 inet6 addr: fe80::226:55ff:fed8:9415/64 Scope:Link UP BROADCAST RUNNING MULTICAST MTU:1500 Metric:1 RX packets:8800501 errors:0 dropped:0 overruns:0 frame:0 TX packets:3286564 errors:0 dropped:0 overruns:0 carrier:0 collisions:0 txqueuelen:1000 RX bytes:12892479880 (12.0 GiB) TX bytes:1184926167 (1.1 GiB) eth1 Link encap:Ethernet HWaddr 00:26:55:D8:94:14 inet addr:192.9.220.1 Bcast:192.9.220.255 Mask:255.255.255.0 inet6 addr: fe80::226:55ff:fed8:9414/64 Scope:Link UP BROADCAST RUNNING MULTICAST MTU:1500 Metric:1 RX packets:11973 errors:0 dropped:0 overruns:0 frame:0 TX pa

Applies to

Models
LightSpeed, VCT
Product
LightSpeed 7.x (VCT, VCT Select, Pro32) V40_H_V64_G_GTM V40_H_V64_G_GTL V40_H_V32_G_GTM V40_H_V32_G_GTL V40_H_H64_G_GTM V40_H_H64_G_GTL V40_H_HS64_G_G Product Name: LightSpeed 7.x (VCT, VCT Select, Pro32) Product photographs Method of remote or software identification - See attachment for a complete list of software commands and typical results. check_config swhwinfo “swhwinfo” will have the following new switches available with the VCT M4 and PF 1.5 M3 software release (software release deploying this “swhwinfo” capability is “06MW03.?”): · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) · -h or –help = Outputs complete details about all the available switches associated with swhwinfo · swhwinfo –c –m –o -da List of Supported Console Types GOC5 Will have a VDARC and VIG GOC6 Will have a VDIG List of softwares the product can run 09MW08.10 09MW08.11 08MW33.2 07MW18.4 06MW03.4 05MW31.6 FAQ's Table weight limit - Up to 500 lb. (227 kg) with ± .25 mm positional accuracy Tube type - D3194T D3193T(green) Current production - M4 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system. Done acording to template
Component
Ethernet cable, CAT5e 1.8 Meter. see long description. Part #: 5329114 Part Type: GP CONSUMABLES Part Weight: 0.00 KGM Repairable: No Returnable: No Hazardous Material: N Description: Ethernet cable, CAT5e 1.8 meter FRU: Yes Status: Orderable Parts Offered as: Current: 5329114 Harvested: Not Available Exchange: Not Available Comments: Per HDVCT Options RSPL, 212008898, 5/4/09.
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.