GE Healthcare · CTLightSpeed260001071500392600040255022919127
LightSpeed 4.X - GRE
Symptom
Scanner Hardware Stopped Scan. Scanner Hardware Stopped Scan
Error message
Error: 260001071 The CCB has reset unexpectedly. The OBC has successfully reconnected to sub-system. See LD for complete error Error: 260001071 The CCB has reset unexpectedly. The OBC has successfully reconnected to sub-system. Fri Feb 10 16:23:44 2006 Error: 260001071 Host: OBC Process: HOP File: hop_node_init.c Line: 347 Function: No System Function Reported Scan: 50039/6/0 Type: None/Unknown Exception Level: Pri/Soft Time: 16:23:44:745 Log Series: 130 The CCB has reset unexpectedly. The OBC has successfully reconnected to sub -system. Reason for reset as reported by CPU: External signal source. CPU reset register: 0x80 Node status: 0x382 State flags: 0x9800682 Error: 260004025 DAS / Collimator firmware object: Lost communications with node applications. See Long Description for full error Thu Jan 13 06:58:23 2005 Error: 260004025 Host: OBC Process: Collimator File: col_cf_in_mail.c Line: 442 Function: No System Function Reported Scan: 50229/1/0 Type: None/Unknown Exception Level: Pri/Most Severe Time: 06:58:23:235 Log Series: 47 Collimator firmware object: Lost communications with node applications. Thu Dec 31 11:45:03 2009 Error:260004025 Host: OBC Process: DAS File: das cf in
Applies to
- Models
- LightSpeed
- Product
- LightSpeed 4.X - GRE See Long Description for detailed information - H2_P_M16 Product Name: LightSpeed 4.X with GRE Product Photographs GE HEALTHCARE Method of remote or software identification - See attachment below for a complete list of software commands and typical results 1. check_config 2 . swhwinfo “swhwinfo” will have the following switches: · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) -h or –help = Outputs complete details about all the available switches associated with swhwinfo swhwinfo –c –m –o -da List of Supported Console types: GOC3 (GRE) GOC4 (GRE) List of software the product can run: 07MW11.10 SP2.0, SP3.0, SP3.1 FAQ's Table weight limit - 180 kg (400 lbs) normal operation - 205 kg (450 lbs) maximum allowed with decreased accuracy Tube type - D3186T D3186T(green) Current production - M4 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Component
- 2159578-3 collimator control board used on 1X/2X/3X/4X/5X See Long description for details Part #: 2159578-3 Part Type: GP CONSUMABLES Part Weight: 2.18 LBR Repairable: No Returnable: No Hazardous Material: N Description: COLLIMATOR CONTROL BOARD FRU: Yes Status: Orderable Comments: Not Available
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.